FDA Medical Device Recalls (openFDA)
Synthes USA HQ, Inc.: Synthes TomoFix Plates The Synthes J5606-C TomoFix Osteotomy System is used for open and closed wedge osteotomies of the medial proximal tibia, lateral proximal tibia, medial and lateral distal femur, treatment of bone and joint deformities, fractures, and malalignment caused by injury or disease such as osteoarthritis.
- Recall number
- Z-0473-2014
- Recalling firm
- Synthes USA HQ, Inc.
- Product
- Synthes TomoFix Plates The Synthes J5606-C TomoFix Osteotomy System is used for open and closed wedge osteotomies of the medial proximal tibia, lateral proximal tibia, medial and lateral distal femur, treatment of bone and joint deformities, fractures, and malalignment caused by injury or disease such as osteoarthritis.
- Status
- Terminated
- Initiated
- 2013-10-09
- Posted
- 2013-12-10
- Terminated
- 2015-12-18
- Reason
- A labeling correction was initiated related to the Surgical Technique Guide J5606-C "TomoFix Osteotomy System" due to the potential for a surgical delay if the surgeon inserts the superior screws at a lesser angle (4 or 6 degrees) and is not aware that those angles are not present.
- Root cause
- Labeling False and Misleading
- Action
- Synthes sent an Urgent Field Safety Notification/Medical Device Labeling Correction letter dated October 9, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to exchange the old surgical technique guide version J5606-C with the new version J5606-D provided with this notification. Review the revised surgical technique guide (Pages 25 & 26). Forward this Field Safety Notification to anyone in their facility that need to be informed. If the surgical technique guide has been forwarded to another facility, contact that facility. Maintain awareness of this notice until all surgical technique guides have been exchanged. Maintain a copy of this notice. Customers with questions were instructed to contact their Synthes Sales Consultant. For questions regarding this recall call 610-719-5000.
- Quantity
- 2912
- Distribution
- Worldwide Distribution - USA (nationwide) Canada and Mexico
- Lot, serial or model codes
- J5606-C
- 510(k) numbers
- ["K023941"]
- PMA numbers
- not on file
- Product code
- KTT
- Firm FEI
- 3005180112
- Firm city
- West Chester
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28