Biomet, Inc.: Green Mamba Suture Passer

Recall number
Z-0472-2016
Recalling firm
Biomet, Inc.
Product
Green Mamba Suture Passer
Status
Terminated
Initiated
2015-12-02
Posted
not on file
Terminated
2016-10-27
Reason
There may be a separation and an insufficient weld between the spring and washer. The weld between the spring and washer may fracture, causing the Mamba Suture Passer instruments jaw to not actuate properly. 6 complaints have been received in which 3 have been reported with a delay up to 10 minutes.
Root cause
Nonconforming Material/Component
Action
On 12/2/2015, URGENT MEDICAL DEVICE RECALL NOTICE notifications were sent to the affected distributors and hospital risk managers via courier with instructions to locate and return the affected products. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Quantity
13
Distribution
FL, NJ, WI, UT, TX, KY, KS, OH, GA, LA, AL, NY, NC, AR, MO, VA, CA, MN, IN, HI
Lot, serial or model codes
Catalog Number: 110010850 Lot Number Identification: 167890, 231120, 253210, 550050, 551580, 716300, 953230
510(k) numbers
not on file
PMA numbers
not on file
Product code
HWQ
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register