FDA Medical Device Recalls (openFDA)
GE Healthcare, LLC: Multi Absorber Original, Disposable M1173310 Multi Absorber Original, Disposable, package of 6 pcs, GE Healthcare Finland Oy Made in US Rx Only The GE Healthcare Multi Absorber Original, Disposable is a disposable product intended for use with the GE Healthcare Advanced Breathing System (later ABS), the GE Healthcare EZchange manifold, the GE Healthcare Compact Block, and the GE Healthcare Compact Block II (later Compact Block). The Multi Absorber should only be used with air, oxygen, nitrous oxide, halothane, enflurane, isoflurane, desflurane and sevoflurane. The device is intended to be used under constant attention of qualified professional healthcare personnel.
- Recall number
- Z-0472-2014
- Recalling firm
- GE Healthcare, LLC
- Product
- Multi Absorber Original, Disposable M1173310 Multi Absorber Original, Disposable, package of 6 pcs, GE Healthcare Finland Oy Made in US Rx Only The GE Healthcare Multi Absorber Original, Disposable is a disposable product intended for use with the GE Healthcare Advanced Breathing System (later ABS), the GE Healthcare EZchange manifold, the GE Healthcare Compact Block, and the GE Healthcare Compact Block II (later Compact Block). The Multi Absorber should only be used with air, oxygen, nitrous oxide, halothane, enflurane, isoflurane, desflurane and sevoflurane. The device is intended to be used under constant attention of qualified professional healthcare personnel.
- Status
- Terminated
- Initiated
- 2013-09-19
- Posted
- 2013-12-13
- Terminated
- 2015-03-10
- Reason
- GE Healthcare has recently become aware of a potential safety issue due to air leakage associated with the CO2 Multi Absorber.
- Root cause
- Device Design
- Action
- GE issued an Urgent Medical Device Correction letter dated September 19, 2013, to all affected customers via Fed Ex overnight mail. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to not use the affected product, isolate all affected product and return the attached form via fax to 800-535-7923. Once the information is received, customers would be contacted with an RMA and replacement information. If product was distributed further the notice should be forwarded to those customers. For questions customers should call 1-800-345-2700, option 2 followed by ooption 2. For foreign customers 1-800-932-0760, option 2. For questions regarding this recall call 262-513-4122. GE issued a press release on 12/30/2013.
- Quantity
- US: 59,721 boxes, 358,326 canisters; OUS: 3,724 boxes, 22,344 canisters
- Distribution
- Worldwide Distribution - USA (nationwide) Puerto Rico and Internationally to Canada, Costa Rica, Mexico, and Venezuela.
- Lot, serial or model codes
- product number M1173310 containing lot numbers 12001 through 13031.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- BSF
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28