Reflexion Medical, Inc.: RefleXion Medical Radiotherapy System - System Label: "*** refleXion REF RXM1000 ***" Physics Guide/Treatment Delivery User Manual/Treatment Planning User Manual: "RefleXion X1 Model: RXM1000"
- Recall number
- Z-0470-2022
- Recalling firm
- Reflexion Medical, Inc.
- Product
- RefleXion Medical Radiotherapy System - System Label: "*** refleXion REF RXM1000 ***" Physics Guide/Treatment Delivery User Manual/Treatment Planning User Manual: "RefleXion X1 Model: RXM1000"
- Status
- Terminated
- Initiated
- 2021-11-24
- Posted
- not on file
- Terminated
- 2024-07-08
- Reason
- Due to dose discrepancy when delivering a plan to a patient in a Non-HFS (Head First Supine) orientation specifically in Feet First Supine (FFS).
- Root cause
- Software design
- Action
- On 11/24/2021 and 12/02/2021, the firm emailed customers to inform them of an ongoing investigation a potential dose discrepancy while treating patients in an orientation other than Head First Supine (HFS), i.e., Feet First Supine (FFS) and Head First Prone (HFP) and Feet First Prone (FFP). On 12/22/2021, the firm sent an "Urgent Safety Notification" letter to customers informing them that the Recalling Firm has confirmed that a dose discrepancy exists for patients treated in any orientation other than Head First Supine (HFS). Customers are instructed, until the updated software is install at their institution, the Recalling Firm recommends using only the HFS orientation when treating patients on the affected systems. It is estimated that a software updated to address the issue will be installed at all sites by mid to late January 2022. For questions, contact the Senior Director of Technical Support at 650-263-8888 or email support@reflexion.com
- Quantity
- 3 devices
- Distribution
- U.S. Nationwide distribution in the states of CA and TX
- Lot, serial or model codes
- Model: RXM1000 Serial Numbers/UDI Codes: X11001 / (01)00860003983805(11)210125(21)X11001 X11002 / (01)00860003983805(11)210324(21)X11002 X11003 / (01)00860003983805(11)210524(21)X11003
- 510(k) numbers
- ["K190978"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 3011716550
- Firm city
- Hayward
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28