FDA Medical Device Recalls (openFDA)
Philips Medical Systems North America Co. Phillips: The product's marketing brochure states: "1.5T coils and accessories...Philips Achieva and Intera 1.5T MR systems". The Identification of the Synergy Flex-M/ Shoulder Coil 1.5T can be found on the driver box. The 12nc of the affected system is labeled as 4522-131-6656x.
- Recall number
- Z-0469-2011
- Recalling firm
- Philips Medical Systems North America Co. Phillips
- Product
- The product's marketing brochure states: "1.5T coils and accessories...Philips Achieva and Intera 1.5T MR systems". The Identification of the Synergy Flex-M/ Shoulder Coil 1.5T can be found on the driver box. The 12nc of the affected system is labeled as 4522-131-6656x.
- Status
- Terminated
- Initiated
- 2009-05-01
- Posted
- 2010-11-26
- Terminated
- 2011-01-06
- Reason
- Combined use of the Synergy Flex-M / Shoulder Coil 1.5T increases the chance of RF interaction and heating up of the coil. This may result in possible burns of the patient.
- Root cause
- Other
- Action
- Philips issued URGENT-Field Safety Notice letters dated May 1, 2009 and October 14, 2009 to their consignees. The letter informed customers that the firm will replace the "old" Flex M/Shoulder Coils with a new Flex-M coil. Consignees could contact Philips Healthcare Call Center at 800-722-9377, #5, #3, #1 and reference FCO 78100290 for any questions.
- Quantity
- 389 units
- Distribution
- Nationwide Distribution: in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VA, WA, WI, WV, WY, and Washington DC. There were no foreign consignees.
- Lot, serial or model codes
- Site Numbers: 10009, 10353, 10409, 10432, 13056, 13057, 13894, 17648, 25958, 25986, 26201, 26219, 26269, 35394, 37986, 38111, 38134, 38143, 38303, 38304, 38420, 38438, 38966, 41345, 41391, 45016, 50336, 59435, 59637, 59712, 62450, 62658, 62832, 63046, 76320, 76321, 76359, 76475, 76847, 82546, 82619, 83080, 83301, 83303, 83306, 83325, 83394, 83396, 83804, 83876, 83940, 83946, 83995, 84433, 84477, 84478, 84700, 84820, 84860, 84887, 84899, 85119, 85139, 85141, 85157, 85158, 85172, 85433, 85442, 85500, 85684, 85685, 85686, 85688, 86019, 86056, 86080, 86085, 86121, 86147, 86184, 86236, 86280, 86281, 86292, 86293, 86313, 86317, 86334, 86335, 86381, 86404, 86406, 86407, 86410, 86557, 86560, 86570, 86595, 86596, 86968, 87011, 87018, 87033, 87038, 87053, 87135, 87145, 87146, 87147, 87148, 87161, 87168, 87183, 87188, 87309, 87310, 87526, 87527, 87605, 87607, 100214, 100459, 100460, 100462, 100469, 100478, 100745, 100785, 100786, 100826, 100832, 100916, 101084, 101111, 101138, 101162, 101165, 101206, 101313, 101347, 101408, 101420, 101429, 101431, 101477, 101481, 101551, 101561, 101584, 101585, 101595, 101608, 101780, 101874, 101877, 101878, 101934, 101946, 101976, 102003, 102004, 102018, 102095, 102110, 102137, 102492, 102547, 102548, 102561, 102569, 102571, 102671, 102709, 102718, 102724, 102889, 102907, 102933, 102975, 102978, 102980, 103041, 103062, 103068, 103099, 103126, 103135, 103136, 103142, 103143, 103149, 103166, 103171, 103207, 103215, 103243, 103247, 103291, 103324, 103333, 103347, 103351, 103417, 103450, 103509, 103530, 103540, 103544, 103547, 103552, 103579, 103580, 103582, 103592, 103611, 103631, 103665, 103688, 103703, 103770, 103782, 103789, 103811, 103847, 103904, 103946, 104022, 104032, 104054, 104069, 104070, 104082, 104095, 104102, 104138, 104163, 104193, 104336, 104340, 104343, 104378, 104392, 104393, 104446, 104467, 104532, 104639, 104765, 104800, 104808, 104871, 104886, 104887, 104897, 104900, 104901, 104912, 104918, 104919, 104942, 105148, 105252, 105271, 105290, 105303, 105314, 105315, 105317, 105348, 105368, 105471, 105486, 105488, 105624, 105657, 105687, 105695, 105741, 105925, 105926, 105929, 105938, 225313, 250050, 250122, 250156, 250293, 250890, 500009, 500034, 504243, 504249, 504342, 504353, 504400, 504403, 504409, 504412, 504527, 504531, 504543, 504608, 504628, 504662, 504670, 504738, 504866, 505021, 505049, 505050, 505275, 505337, 505357, 505490, 505514, 505568, 505678, 505777, 505785, 505914, 505940, 505941, 505943, 506034, 506044, 506188, 506190, 506203, 506219, 506296, 506324, 506389, 506570, 506691, 506777, 519124, 519127, 519129, 519147, 519216, 519522, 519525, 519545, 520002, 520113, 520354, 520709, 521429, 530363, 530651, 531767, 533741, 536318, 536649, 542669, 544006, 544503, 545940, 546669, 547235, 550391, 556698, 557719, 557722, 557723, 557724, 557791, 557793, 557801, 557805, 557806, 557807, 557813, 557814, 557843, 557844, 557904, 558118, 558147, 558177, 558200, 558369, 558371, 558374, 558381, 558382, 41923351, and 43668550.
- 510(k) numbers
- ["K001796","K031815"]
- PMA numbers
- not on file
- Product code
- LNH
- Firm FEI
- 1000524572
- Firm city
- Bothell
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28