Bard Peripheral Vascular Inc: Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210; BARD MARQUEE 12G X 10 CM, REF: MQ1210; BARD MARQUEE Kit 12G X 13 CM, REF: MQK1213; BARD MARQUEE 12G X 13 CM, REF: MQ1213

Recall number
Z-0467-2025
Recalling firm
Bard Peripheral Vascular Inc
Product
Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210; BARD MARQUEE 12G X 10 CM, REF: MQ1210; BARD MARQUEE Kit 12G X 13 CM, REF: MQK1213; BARD MARQUEE 12G X 13 CM, REF: MQ1213
Status
Open, Classified
Initiated
2024-09-30
Posted
2024-11-15
Terminated
not on file
Reason
Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could lead to needle detachment and misplacement, which could lead to tissue damage and bleeding.
Root cause
Process change control
Action
On 9/30/24, removal notices were sent to customers who were asked to do the following: 1) Discontinue use of product listed in the letter, and immediately dispose of all affected product remaining in your possession in accordance with your local facility s process. 2) Share this notification with all users within your facility network of the product to ensure they are also aware of this field action. If the affected products have been distributed further, please share this notice to ensure that these entities are contacted and informed of the recall. 3) Complete and return the response form via email to BDRC27@BD.COM - If you purchased this product from a distributor, contact your distributor for further instructions and resolution. - Affected lots distributed from 05 April 2023 through 23 Aug 2024. If further assistance is required contact the firm: - North American Regional Complaint Center: 1-844-823-5433, Mon - Fri 8:00am and 5:00pm CT, productcomplaints@bd.com - Recall Related Questions: BDRC27@BD.COM On 1/15/2025, additional removal notices were mailed or emailed to distributors informing them of additional affected devices.
Quantity
71,498
Distribution
Worldwide - US Nationwide including in the states of WA, LA, MN, PA, UT, SC, OK, KY, AZ, IL, TX, NJ, NY, FL, TN, IN, OH, NC, CO, VA, GA, MD, CA, AL, WI, MO, IA, CT, NE, MI, MS, NV, KS, OR, SD, ID, NH, RI, WY, AR, MA, DE, ND, NM, AK, PR and the countries of Portugal, Spain, South Africa, Greece, United Arab Emirates, United Kingdom, Sweden, Finland, Belgium, Korea, Australia, Canada.
Lot, serial or model codes
REF/UDI-DI/Lot(Expiration): MQK1210/00801741097263/1490758(9/28/2025), 1506044(1/28/2026), 1506045(1/28/2026), 1511827(2/28/2026), 1514692(3/28/2026), 1518977(4/28/2026), 1518978(4/28/2026), 1518979(4/28/2026), 1519143(4/28/2026), 1520507(5/28/2026), 1520508(5/28/2026), 1522157(5/28/2026), 1522158(5/28/2026), 1522577(5/28/2026), 1523785(6/28/2026), 1523914(6/28/2026), 1523915(6/28/2026), 1525370(6/28/2026), 1525371(6/28/2026), 1525373(6/28/2026), 1528417(7/28/2026), 1528418(7/28/2026), 1528495(7/28/2026), 1528496(7/28/2026), 1528497(7/28/2026), 1528560(7/28/2026), 0001618867(3/28/2028), 0001620281(3/28/2028), 0001632675(6/28/2028); MQ1210/00801741097119/1514460(3/28/2026), 1514461(3/28/2026), 1519144(4/28/2026), 1519145(4/28/2026), 1519456(4/28/2026), 1520066(5/28/2026), 1520067(5/28/2026), 1520506(5/28/2026), 1523783(6/28/2026), 1523784(6/28/2026), 0001618258(3/28/2028), 0001631611(6/28/2028), 0001633216(7/28/2028); MQK1213/00801741097270/1543517(10/28/2026), 1559312(2/28/2027), 1562724(3/28/2027), 1569737(4/28/2027), 1581652(7/28/2027); MQ1213/00801741097126/1558191(2/28/2027), 1559313(2/28/2027)
510(k) numbers
["K133948"]
PMA numbers
not on file
Product code
KNW
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register