Philips And Neusoft Medical Systems Co., Ltd.: NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.

Recall number
Z-0466-2013
Recalling firm
Philips And Neusoft Medical Systems Co., Ltd.
Product
NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.
Status
Terminated
Initiated
2012-10-18
Posted
2012-12-04
Terminated
2013-07-08
Reason
The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.
Root cause
Device Design
Action
Philips and Neusoft Medical Systems Co., Ltd. sent an Urgent Device Correction NeuViz Dual, NeuViz 16 letter, dated Oct. 18, 2012, to all affected US customers and distributors. The letter described the product, problem, and actions to be taken by the customers. Customers are advised to: 1. Have a mechanic check the patient table top screws fixing. Detailed instructions are attached as Annex with the letter. 2. If it is found that table top screws are improperly assembled, stop using the product and contact local Service personnel for repair. Philips and Neusoft Medical Systems Co., Ltd is in the process of upgrading devices with aforementioned kits and will contact customers to arrange for an update to affected systems. If you have any questions about this corrective action program, please contact the Service Support Department, Neusoft Medical Systems Co. Ltd. at nms-service@neusoft.com For questions regarding this recall call 425-487-7665.
Quantity
20 units of NeuViz 16 CT Scanner Systems and 25 units of NeuViz Dual series
Distribution
Nationwide Distribution including North Carolina, Ohio, Texas, and Puerto Rico.
Lot, serial or model codes
NeuViz 16 system Serial numbers:  N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120022, N16E120020, N16E120023, N16E120024.   Neuviz Dual Serial numbers:  400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR090003, NDHR090012, and NDHR090013.
510(k) numbers
["K062451","K071308","K090173","K092742"]
PMA numbers
not on file
Product code
JAK
Firm FEI
3003575350
Firm city
Shenyang
Firm state
Liaoning
Firm country
China

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register