Skip to the content

FDA Medical Device Recalls (openFDA)

AGA Medical Corporation: AMPLATZER Duct Occluder, Order No. 9-PDA-007, Device Size 12/10mm, Length 8 mm. Sterile EO. Single Use Only. Manufactured by AGA Medical Corporation, 682 Mendelssohn Avenue, Golden Valley, MN 55427 USA. The AMPLATZER Duct Occluder is a self-expandable device made from a Nitinol wire mesh. A retention skirt on the aortic side provides secure positioning in the ampulla of the ductus. As the occluder is implanted, it expands outward and the wires push against the wall of the ductus. Polyester fabric is sewn into the occluder with polyester thread. The fabric induces thrombosis that closes the communication.

Recall number
Z-0465-06
Recalling firm
AGA Medical Corporation
Product
AMPLATZER Duct Occluder, Order No. 9-PDA-007, Device Size 12/10mm, Length 8 mm. Sterile EO. Single Use Only. Manufactured by AGA Medical Corporation, 682 Mendelssohn Avenue, Golden Valley, MN 55427 USA. The AMPLATZER Duct Occluder is a self-expandable device made from a Nitinol wire mesh. A retention skirt on the aortic side provides secure positioning in the ampulla of the ductus. As the occluder is implanted, it expands outward and the wires push against the wall of the ductus. Polyester fabric is sewn into the occluder with polyester thread. The fabric induces thrombosis that closes the communication.
Status
Terminated
Initiated
2005-09-06
Posted
2006-01-31
Terminated
2006-12-26
Reason
The pointed end of a screw securing the occluder to the delivery cable could scrape off microscopic fragments of the polytetrafluorethylene (PTFE) plastic lining of the sheath. Those shavings could potentially migrate into a patient's bloodstream.
Root cause
Other
Action
User facilities and distributors were notified via email beginning 09/06/05. Additionally a hard copy of the recall letter was sent via certified mail or Federal Express. AGA requests that the Response Form be completed and faxed or emailed back to them. AGA Medical staff will assist users in returning product distributed within the US.
Quantity
18,725 total
Distribution
OUS to the following countries: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Denmark, El Salvador, Finland, France, Germany, Greece, Guatemala, Hungary, India, Indonesia, Isreal, Italy, Jordan, Korea, Kuwait, Malaysia, Mexico, Morocco, Netherlands, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Saudia Arabia, Singapore, Slovakia, South Africa, Spain, Sweeden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uruguay, Venezuela, Viet Nam and United Kingdom.
Lot, serial or model codes
all lots
510(k) numbers
not on file
PMA numbers
not on file
Product code
MAE
Firm FEI
3005334138
Firm city
Golden Valley
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register