Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF): Instructions for Use for all model numbers of the Medtronic Abre Venous Self-Expanding Stent System.

Recall number
Z-0464-2022
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Product
Instructions for Use for all model numbers of the Medtronic Abre Venous Self-Expanding Stent System.
Status
Terminated
Initiated
2021-11-19
Posted
not on file
Terminated
2024-08-16
Reason
There have been reports of stent migration.
Root cause
Other
Action
The firm issued letters dated November 2021 on 11/19/2021 via 2-day UPS delivery to the U.S. consignees. The letter informs the consignee of the upcoming updates to the Instruction for Use (IFU) for the Abre venous self-expanding stent system and explains the updates will provide new information to help mitigate the risk of possible stent migration. The Customer Instructions in the letter requests the enclosed upcoming updates to the IFU be reviewed, the letter is shared with all those who need to be aware within their organization, patients should continue to be monitored per their practice's normal follow-up procedures, and the enclosed Customer Confirmation Form be completed and emailed to the firm. The updated electronic IFU will be available before the end of the calendar year. In Outside the United States (OUS) geographies, communication of the Medical Device Correction letter began 11/24/2021 via regionally approved methods. Consignees will be notified of the upcoming IFU change, as well as provided the content of the IFU change. Consignees will be asked to complete and return a confirmation certificate confirming they have received the notification. The IFU for OUS consignees will be available in the summer of 2022.
Quantity
21,653 devices
Distribution
Distribution was nationwide. There was also military/government distribution. Foreign distribution was made to Argentina, Australia, Austria, Bahrain, Belgium, Botswana, Bulgaria, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Jordan, Kenya, Lithuania, Malaysia, Republic of Moldova, Netherlands, New Zealand, Nicaragua, Norway, Oman, Peru, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, and Viet Nam,
Lot, serial or model codes
The IFU for all model numbers and serial numbers.
510(k) numbers
not on file
PMA numbers
["P200026"]
Product code
QAN
Firm FEI
2182208
Firm city
Mounds View
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register