Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corp: IntellaTip MiFi XP Asymmetric (N4) Curve 7/110/2.5/8-8 Material Number: M004PM4500N40; Catalog number: PM4500N4; Cardiac: The IntellaTip MiFi XP Catheter is indicated for interruption of accessory atrioventricular conduction pathways associated with tachycardia, for treatment of AV nodal reentrant tachycardia, for treatment of atrial flutter tachycardia, and for creation of complete AV block in patients with a rapid ventricular response to an atrial arrhythmia typically chronic, drug refractory atrial fibrillation.

Recall number
Z-0464-2016
Recalling firm
Boston Scientific Corp
Product
IntellaTip MiFi XP Asymmetric (N4) Curve 7/110/2.5/8-8 Material Number: M004PM4500N40; Catalog number: PM4500N4; Cardiac: The IntellaTip MiFi XP Catheter is indicated for interruption of accessory atrioventricular conduction pathways associated with tachycardia, for treatment of AV nodal reentrant tachycardia, for treatment of atrial flutter tachycardia, and for creation of complete AV block in patients with a rapid ventricular response to an atrial arrhythmia typically chronic, drug refractory atrial fibrillation.
Status
Terminated
Initiated
2015-11-20
Posted
not on file
Terminated
2019-11-27
Reason
During routine monitoring of complaints, BSC noted that the IntellaTip MiFi XP N4 complaint rate for distal tip bends/kinks was above the anticipated rate in BSCs product risk documentation.
Root cause
Device Design
Action
Boston Scientific sent an Urgent Medical Device Removal letters to all affected customers dated November 20, 2015 via Fed Ex or other equivalent service. The letter identified the affected products, problem and actions to be taken. Customers were instructed to contact their local Sales Representative to answer any questions.
Quantity
1556 total - all models
Distribution
Worldwide Distribution - US Nationwide in the state of: PA, FL, IN, MI, TN, AL, IL, KY, MI, TX, OH, MO, MA, MN, ID, AL, TN, VA, CA and internationally to Austria, Australia, Belgium, Switzerland, Germany, Great Britain, Greece, Hong Kong, Israel, Iran, Italy, Japan, Kuwait, Latvia, Malaysia, Netherlands, Saudi Arabia, Singapore, Thailand, and South Africa.
Lot, serial or model codes
Material Number: M004PM4500N40;  Catalog number: PM4500N4; Lot numbers: 16309584, 16615976, 16623754, 16623755, 16722434, 16736927, 16739588, 16743271, 16872124, 16999075, 17118111, 17157418, 17201805, 17222051, 17222052, 17370483, 17376573, 17376574, 17436562, 17436563, 17436565, 17451802, 17478950, 17499145, 17499152, 17590503, 17611785, 17625992, 17768306, 17791420, 17925492, 17925493, 18012208, 18161715. Expiration Date Range:8/15/2016 to 7/2/2018.
510(k) numbers
not on file
PMA numbers
["P020025S036"]
Product code
OAD
Firm FEI
3001236349
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register