FDA Medical Device Recalls (openFDA)
Zimmer, Inc.: NexGen Complete Knee Solution CR Articular Surface Provisional-Yellow 9mm Height Nonsterile Devices are used in total knee arthroplasty and indicated for patients with severe pain and disability due to conditions such as arthritis, collagen disorders, post-traumatic loss of joint configuration, flexion deformities, and in revision of previous surgical attempts.
- Recall number
- Z-0462-2014
- Recalling firm
- Zimmer, Inc.
- Product
- NexGen Complete Knee Solution CR Articular Surface Provisional-Yellow 9mm Height Nonsterile Devices are used in total knee arthroplasty and indicated for patients with severe pain and disability due to conditions such as arthritis, collagen disorders, post-traumatic loss of joint configuration, flexion deformities, and in revision of previous surgical attempts.
- Status
- Terminated
- Initiated
- 2013-11-19
- Posted
- 2013-12-06
- Terminated
- 2015-05-06
- Reason
- This is an expansion of the June 2013 recall of NexGen Micro components. Five (5) additional instruments have been identified that are associated only with the NexGen Micro components. There is the potential for these devices to be inadvertently utilized during surgery, resulting in a delay. The associated implants would not be available due to the previous recall.
- Root cause
- Labeling design
- Action
- Zimmer sent an Urgent Medical Device Recall Notice dated November 19, 2013, to all affected customers. The Notice outlined the products recalled; the reason for the recall, possible adverse events related to the recall and directed consignees to locate, quarantine and count all affected product recording the information on the Inventory Return Form acknowledging that recall directions were followed. Further questions should be directed to 1-877-946-2761. Affected product will be removed by a Zimmer Sales representative, Question should be directed to 1-877-946-2761 8a.m - 5 p.m. EST. For questions regarding this recall call 800-613-6131.
- Quantity
- 684 units
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Australia, Belgium, Canada, Denmark, Ecuador, Germany, India, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, Peru, Singapore, South Africa, Spain, Switzerland, UAE, and UK.
- Lot, serial or model codes
- Part Number 00-5971-031-09
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HWT
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28