FDA Medical Device Recalls (openFDA)
Novoste Corporation: Novoste, Beta-Cath 3.5F System, New B-Rail 3.5F Delivery Catheter, with Indicator of Source Train (IST), Novoste Corporate, 3890 Steve Reynolds Blvd, Norcross, GA 30093, Tel: +1770-717-0904, Fax: +1770-717-1283, 1-800-NOVOSTE (1-800-668-6783), www.novoste.com.
- Recall number
- Z-0462-03
- Recalling firm
- Novoste Corporation
- Product
- Novoste, Beta-Cath 3.5F System, New B-Rail 3.5F Delivery Catheter, with Indicator of Source Train (IST), Novoste Corporate, 3890 Steve Reynolds Blvd, Norcross, GA 30093, Tel: +1770-717-0904, Fax: +1770-717-1283, 1-800-NOVOSTE (1-800-668-6783), www.novoste.com.
- Status
- Terminated
- Initiated
- 2002-01-25
- Posted
- 2003-01-30
- Terminated
- 2003-11-17
- Reason
- Mislabeling. Source calibrated incorrectly, resultant dose rate lower than expected.
- Root cause
- Other
- Action
- Customers were notified via Advisory Letter followed up with a Recall Letter and Customer Informational Letter on/about January 24, 2002. Customers were advised to discontinue use of the product and return the affected product to Novoste Corp. to be recalibrated to obtain a correct calibration certificate.
- Quantity
- 50 units
- Distribution
- Nationwide. (AR, AZ, CA, FL, GA, IA, ID, IL, KY, MD, MI, MN, MO, NC, NM, NV, NY, OH, PA, TN, TX, VA, WA). No foreign or government accounts involved.
- Lot, serial or model codes
- Catalog Numbers: TDA-0030 - Serial Numbers: 81431, 81761, 82548, 83007, 83926, 84395, 85014, 85021, 86333, 86489, 86874, 86889, 86897, 87097, 87103, 87189, 88592, 88692, 88734, 88738, 89035, 89037, 89058, TDA-0040- Serial Numbers: 77252, 81012, 83214, 85327, 87816, 87834, 88591, 88598, 88617, 88622, 88626, 88632, 88681, 88697, 88691, 88697, 88702, 88712, 88718, 88719, 88720, 88721, 88736, 88739, 89045, 89049, 89054.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JAJ
- Firm FEI
- 3000205626
- Firm city
- Norcross
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28