Skip to the content

FDA Medical Device Recalls (openFDA)

Wilson-Cook Medical Inc.: QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-4X3, Order Number G22654; used to dilate strictures of the biliary tree

Recall number
Z-0459-2024
Recalling firm
Wilson-Cook Medical Inc.
Product
QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-4X3, Order Number G22654; used to dilate strictures of the biliary tree
Status
Open, Classified
Initiated
2023-11-03
Posted
2023-12-01
Terminated
not on file
Reason
Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally.
Root cause
Under Investigation by firm
Action
Cook Medical issued an URGENT: MEDICAL DEVICE REMOVAL notice on 11/03/2023 by mail. The notice explained the problem with the device, risk to health, and requested the following actions be taken: Customer: Cease use, quarantine, and return all affected devices. If devices were further distributed, those to whom the devices were sent must be notified. Distributor: Quarantine all affected devices, notify your customers to whom you distributed devices. Return all devices to Cook Medical.
Quantity
553 devices
Distribution
US
Lot, serial or model codes
UDI/DI 00827002226548 (each), 10827002226545 (case), Lot Numbers: W4720848, W4721792, W4721795, W4722159, W4722981, W4723304, W4723846, W4724692, W4737576, W4738717, W4742525, W4743193, W4743605, W4743606, W4743607, W4743608, W4744022, W4744441, W4748043, W4748946, W4749748, W4750264, W4750265, W4750910, W4751772, W4751906, W4751914, W4753360, W4753361, W4753362, W4753365, W4753816, W4755138, W4756901, W4756903, W4756905, W4757732, W4758647, W4759511, W4759976, W4760715, W4761738, W4761740, W4761935, W4762150
510(k) numbers
["K171223"]
PMA numbers
not on file
Product code
FGE
Firm FEI
1037905
Firm city
Winston Salem
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register