Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics Inc.: Reagent Management System (RMS) Dimension EXL with LM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension clinical chemistry system. The RMS doubles the available storage of refrigerated Flex reagent cartridges on the Dimension system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex reagent cartridges, allowing the Dimension system to process samples without interruption.

Recall number
Z-0458-2016
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Reagent Management System (RMS) Dimension EXL with LM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension clinical chemistry system. The RMS doubles the available storage of refrigerated Flex reagent cartridges on the Dimension system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex reagent cartridges, allowing the Dimension system to process samples without interruption.
Status
Terminated
Initiated
2015-11-10
Posted
2015-12-22
Terminated
2017-08-23
Reason
Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container.
Root cause
Nonconforming Material/Component
Action
Siemens sent an Urgent Medical Device Correction (UMDC # DI-16-01.A.US) and Urgent Field Safety Notice (UFSN # DI-16-01.A.OUS) letters dated November 2015 to all affected customers. The letters identified the affected product, problem and actions to be taken. For questions contact your local Siemens technical support representative.
Quantity
106
Distribution
US Nationwide Distribution: FL, NY, SC, TX, LA, AZ, WI, IA, IL, OH, AL, AK, NH, OK, MD, CA, NM, MS, KY, NJ, MN, GA, CO, MA, SD, TN, MO, PA, MT, SC, NE, KS, VA, AL, NV, WY, OR, PR, MO, OK, GU, HI, AR, ME, VT, CT, NC and the countries of Canada and Mexico.
Lot, serial or model codes
Dimension EXL with LM Catalog Numbers 778021.911 (Domestic)/ 778021.931 (Intl). All serial numbers.
510(k) numbers
["K130276"]
PMA numbers
not on file
Product code
JJE
Firm FEI
3002755490
Firm city
Brookfield
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register