FDA Medical Device Recalls (openFDA)
Nonin Medical, Inc: Nonin Model 7600B cerebral oximeter pod. Used with the Nonin Model 7600 Regional Oximeter with Equanox Technology and Bluetooth Wireless Technology. Nonin Medical, Inc., 13700 First Avenue North, Plymouth, Minnesota 55441-5443 USA.
- Recall number
- Z-0458-2010
- Recalling firm
- Nonin Medical, Inc
- Product
- Nonin Model 7600B cerebral oximeter pod. Used with the Nonin Model 7600 Regional Oximeter with Equanox Technology and Bluetooth Wireless Technology. Nonin Medical, Inc., 13700 First Avenue North, Plymouth, Minnesota 55441-5443 USA.
- Status
- Terminated
- Initiated
- 2009-10-23
- Posted
- 2009-11-25
- Terminated
- 2012-03-29
- Reason
- A malfunction was detected with one of the components that, under rare circumstances, may result in the overheating of the POD. This overheating, while unlikely to occur, has the potential to cause a burn if the POD is in direct contact with the patient.
- Root cause
- Device Design
- Action
- An Urgent: Medical Device Recall letter was emailed to customers on 10/23/2009. The letter described the issue and hazard and asked that the use of the device be discontinued and to immediately contact Nonin for return and repair/replacement instructions. Questions or concerns should be directed to John Dalpee at 763-577-3166.
- Quantity
- 62
- Distribution
- Worldwide Distribution -- USA, including states of LA, MN, NC, and NY and countries of Canada, Germany, Italy, Netherlands, and United Kingdom.
- Lot, serial or model codes
- Serial numbers: 500814470, 500814472, 500814474, 500814478, 500814480, 500814484, 500823139, 500823145, 500823146, 500823154, 500823159, 500823161, 500823164, 500823166, 500823172, 500823174, 500823175, 500823176, 500823181, 500823182, 500849316, 500849318, 500849321, 500849325, 500849331, 500849333, 500849336, 500849337, 500849338, 500849342, 500849343, 500849344, 500849347, 500849348, 500849349, 500849350, 500849351, 500849353, 500849356, 500849357, 500849358, 500849360, 500849361, 500849366, 500849367, 500849368, 500849369, 500849370, 500849372, 500849373, 500849375, 500849378, 500849380, 500849386, 500849387, 500849388, 500849389, 500849390, 500857060, 500857061, 500857064, and 500874926.
- 510(k) numbers
- ["K090807"]
- PMA numbers
- not on file
- Product code
- MUD
- Firm FEI
- 2183646
- Firm city
- Plymouth
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28