Becton Dickinson & Co.: BD ProbeTec ET Urine Processing Kit, 4x25, catalog 440454, packaged in a carton containing 4 storage bags identified as BD ProbeTec ET Urine Processing Pouches, 25 pouches/bag. Carton and bags labeled in part ***Becton, Dickinson and Company, 7 Loveton Circle, Sparks, MD 21152***

Recall number
Z-0455-05
Recalling firm
Becton Dickinson & Co.
Product
BD ProbeTec ET Urine Processing Kit, 4x25, catalog 440454, packaged in a carton containing 4 storage bags identified as BD ProbeTec ET Urine Processing Pouches, 25 pouches/bag. Carton and bags labeled in part ***Becton, Dickinson and Company, 7 Loveton Circle, Sparks, MD 21152***
Status
Terminated
Initiated
2005-01-10
Posted
2005-02-03
Terminated
2005-04-19
Reason
An in-vitro diagnostic kit reagent may cause false negative clinical results in patients tested for sexually transmitted infections.
Root cause
Other
Action
BD notified consignees by phone on 01/10/05 and advised that users transporting patient urine samples may experience increased indeterminate results or false negative results when testing the samples on the BD ProbeTec ET CT/GC Amplified DNA Assay. Subsequest to the phone notification, kit users identified as utilizing the Urine Processing Pouches were further notified by a letter. The letter describes 4 criteria contributing to the risk associated with false negative and indeterminate results. The letter advised users to discontinue use of the kits and discard for replacement. BD further advised in the notification that if the user is not utilizing the Amplification Control (AC), the firm recommends that they evaluate if urine specimens were transported and/or stored at refrigerated temperatures in the kit''s urine pouches. If specimens were transported and/or stored under the noted criteria, BD recommends that users should notify physicians who received reports of negative patient results and indicate that there is a possibility that these results were incorrect. BD plans to provide a 'Dear Physician' letter that would facilitate that notification upon request of the users.
Quantity
2175 kits
Distribution
Product was distributed to direct accounts in the US and to affiliated distributors internationally.
Lot, serial or model codes
Lot/Exp 4103719 October 31, 2005 4118463 October 31, 2005 4119670 November 30, 2005 4124224 November 30, 2005 4133122 November 30, 2005 4105126 October 31, 2005 4118129 November 30, 2005 4104256 October 31, 2005 4091913 October 31, 2005 4104254 October 31, 2005 4090438 October 31, 2005 4090259 October 31, 2005 4084784 October 31, 2005 4014879 July 31, 2005 4016963 July 31, 2005 4023837 July 31, 2005 4023840 August 31, 2005 4041551 August 31, 2005 4083445 October 31, 2005 4133902 November 30, 2005
510(k) numbers
not on file
PMA numbers
not on file
Product code
LSL
Firm FEI
1119779
Firm city
Sparks
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register