Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: Zimmer UNIVERSAL Power System Loaner and Modular Electric/Battery Single Trigger Handpiece. Rx only Made in Switzerland

Recall number
Z-0453-2016
Recalling firm
Zimmer, Inc.
Product
Zimmer UNIVERSAL Power System Loaner and Modular Electric/Battery Single Trigger Handpiece. Rx only Made in Switzerland
Status
Terminated
Initiated
2015-11-10
Posted
2015-12-21
Terminated
2016-02-19
Reason
The firm identified a malfunction which made it possible for a handpiece to start by itself when the power source is connected.
Root cause
Device Design
Action
Zimmer sent an Urgent Medical Device Correction letter dated November 9, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1. Review this notification. 2. You will receive copies of notification sent directly to hospital risk managers and surgeons in your territory. As necessary, review and facilitate understanding of this notification by those entities. 3. Further identify hospital risk managers and/or surgeons in your territory who should also receive notification of this update, and supply the information to the entities you have identified. 4. If after reviewing this notification you have further questions or concerns, please call 330-364-0989 between the hours of 8 a.m. and 5 p.m. EST, Monday through Friday. Zimmer instructed their direct customers that they will upgrade the device with the new electronic during their annual preventative maintenance activities. Until the new upgrade is installed, the firm instructed their customers to follow the Instructions for Use.
Quantity
89 units
Distribution
US Distribution to the states of : AR, CA, CO, CT, FL, GA, HI, IL, KS, KY, LA, ME, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, PA, RI, SC, TX, UT, VA, WA, WI, WV and WY.
Lot, serial or model codes
Model # 01-8507-400-10 & Model # 89-8507-400-10
510(k) numbers
not on file
PMA numbers
not on file
Product code
HAB
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register