FDA Medical Device Recalls (openFDA)

OmniGuide, Inc.: OmniGuide BP-ROBOTIC ( BeamPath Robotic), Single Use Cat. # BP-ROBOTIC The OmniGuide Fiber is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues including intra-oral tissues. It is indicated in the medical specialties of general and plastic surgery, oral/maxillofacial surgery, dentistry, dermatology, gynecology, otorhinolaryngology, gastroenterology, neurosurgery, urology, and pulmonology, and can be used in open surgical procedures as well as endoscopic minimally invasive procedures in conjunction with rigid or flexible endoscopes, such as in laryngoscopy, gastroscopy, colonoscopy, laparoscopy, thoracoscopy, hysteroscopy and bronchoscopy

Recall number
Z-0453-2013
Recalling firm
OmniGuide, Inc.
Product
OmniGuide BP-ROBOTIC ( BeamPath Robotic), Single Use Cat. # BP-ROBOTIC The OmniGuide Fiber is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues including intra-oral tissues. It is indicated in the medical specialties of general and plastic surgery, oral/maxillofacial surgery, dentistry, dermatology, gynecology, otorhinolaryngology, gastroenterology, neurosurgery, urology, and pulmonology, and can be used in open surgical procedures as well as endoscopic minimally invasive procedures in conjunction with rigid or flexible endoscopes, such as in laryngoscopy, gastroscopy, colonoscopy, laparoscopy, thoracoscopy, hysteroscopy and bronchoscopy
Status
Terminated
Initiated
2012-11-05
Posted
2012-12-03
Terminated
2014-10-16
Reason
Failure of glue joint, allowed a portion of the single use device to shift inside, may separate and fall into the patient
Root cause
Process control
Action
OmniGuide issued letter notifiction on November 5, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that an OmniGuide sales representative would contact them immediately to recover the affected product. For any questions customers were instructed to call 617-551-8561. For questions regarding this recall call 617-551-8410.
Quantity
45 units
Distribution
Nationwide Distribution including AZ, FL, MA, NY, OH, and OR.
Lot, serial or model codes
Lot Numbers: LA121003AO-P1 LA121003AQ-P1 LA121001CB-P1 LA121003AL-P1 LA121002AC-P1 LA121001CC-P1 LA121001BZ-P1 LA121001BZ-P1 LA121001AR-P1 LA121001AQ-P1 LA121001CA-P1 LA120926BA-P1 LA120926AS-P1 LA120919BG-P1 LA120920AF-P1 LA120924AG-P1        :
510(k) numbers
["K070157"]
PMA numbers
not on file
Product code
GEX
Firm FEI
3005350457
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register