Synthes USA HQ, Inc.: Titanium Trochanteric Fixation Nail (TFN)System Intended to treat stable and unstable fractures of the proximal femur.

Recall number
Z-0452-2014
Recalling firm
Synthes USA HQ, Inc.
Product
Titanium Trochanteric Fixation Nail (TFN)System Intended to treat stable and unstable fractures of the proximal femur.
Status
Terminated
Initiated
2013-07-30
Posted
2013-12-05
Terminated
2015-09-10
Reason
Two lots of Titanium Trochanteric Fixation Nail (TFN) System are missing the second bend in the nail.
Root cause
Nonconforming Material/Component
Action
Synthes sent an Urgent Notice: Medical Device Recall letter dated July 30, 2013. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to call Synthes at 1-800-479-6329 if they have any of the affeccted product to obtain a Return Authorization Number, Complete the Verification Section of the letter by checking the appropriate box indicating the affected product has been located. Return the Verification Form with the product to: Credit/Returns Synthes 1101 Synthes Avenue Monument, CO 80132. Customers should complete and return the Verification Frorm even if they do not have any of the affected product and return by fax to 1-888-386-2077 or Scan/email: FieldAction@synthes.com. Customers with questions were instructed to call 610-719-5450 or contact their Synthes Trauma Sales Consultant. For questions regarding this recall call 610-719-5000.
Quantity
17
Distribution
Nationwide Distribution including IL, CA, AZ, FL, TX, UT, VA, NY, NM, and MT.
Lot, serial or model codes
Part # 456.391S with lot numbers 5536930, exp. 5/20/2016, and 7161401, exp. 12/31/2021.
510(k) numbers
["K011857"]
PMA numbers
not on file
Product code
HSB
Firm FEI
3005180112
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register