Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho Clinical Diagnostics Inc: VITROS Immunodiagnostic Products-SARS-CoV-2 IgG Calibrator - Product Usage: In-Vitro Diagnostic (EUA) an aid in identifying individuals with an adaptive immune response to SARS-CoV-2, indicating recent or prior infection. Emergency use of this test is limited to authorized laboratories-- Product Code (Unique Identifier): 6199920 (10758750033362)

Recall number
Z-0450-2021
Recalling firm
Ortho Clinical Diagnostics Inc
Product
VITROS Immunodiagnostic Products-SARS-CoV-2 IgG Calibrator - Product Usage: In-Vitro Diagnostic (EUA) an aid in identifying individuals with an adaptive immune response to SARS-CoV-2, indicating recent or prior infection. Emergency use of this test is limited to authorized laboratories-- Product Code (Unique Identifier): 6199920 (10758750033362)
Status
Terminated
Initiated
2020-09-30
Posted
not on file
Terminated
2023-12-07
Reason
Elevated sample results signal/cutoff (s/c) values when using eight (8) lots of VITROS immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator and decreased signal in one (1) lot of VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Control
Root cause
Nonconforming Material/Component
Action
Ortho Clinical Diagnostics issued Urgent Product Correction Notice on 30 Sept 2020 via(CL2020-246) of the issue and instructed as follows: If they are unable to calibrate (now or in the future) using an affected lot of VITROS Anti-SARS-CoV-2 IgG Calibrator Ortho will replace their remaining inventory as indicated on their Confirmation of Receipt form. NOTE: If they have a successful calibration, and quality control results are acceptable, the laboratory may continue using the affected lots for qualitative detection of IgG antibodies to SARS-CoV-2. If they are using the affected product to qualify units as high titer for CPT, consult with your medical director to determine if they should continue testing while waiting for their replacement product. Ortho will replace or credit their remaining inventory as indicated on their Confirmation of Receipt form. Post this notification by each system that processes the affected lots of VITROS Anti-SARS-CoV-2 IgG Calibrator or Control. Complete the enclosed Confirmation of Receipt form no later than October 9, 2020. Indicate the quantity of their remaining inventory to be replaced or credited. Please forward this notification if the product was distributed outside of their facility
Quantity
651 units
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Australia, Brazil, Chile, China, Colombia, India, Japan, Singapore 609917, United Kingdom, Italy.
Lot, serial or model codes
Lot Number/ Exp. Date: 0165 24-NOV-20 0180 19-NOV-20 0185 27-NOV-20 0190 21-NOV-20 0200 23-NOV-20 0205 27-NOV-20 0220 03-DEC-20 0230 03-DEC-20
510(k) numbers
not on file
PMA numbers
not on file
Product code
QMN
Firm FEI
2250051
Firm city
Raritan
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register