Skip to the content

FDA Medical Device Recalls (openFDA)

CareFusion 303, Inc.: SmartSite Low Sorbing Infusion Set with Texium Close Male Luer, Model No. 24301-0007T The SmartSite Low Sorbing Infusion Sets with Texium Male Luer are used to administer fluid and medications through a needle or catheter inserted into the patient's artery or vein.

Recall number
Z-0450-2014
Recalling firm
CareFusion 303, Inc.
Product
SmartSite Low Sorbing Infusion Set with Texium Close Male Luer, Model No. 24301-0007T The SmartSite Low Sorbing Infusion Sets with Texium Male Luer are used to administer fluid and medications through a needle or catheter inserted into the patient's artery or vein.
Status
Terminated
Initiated
2013-11-08
Posted
2013-12-04
Terminated
2014-05-27
Reason
CareFusion is recalling the SmartSite Low Sorbing Infusion Set with Texium Closed Male Luer because it is not assembled with low-sorbing tubing.
Root cause
Employee error
Action
The firm, CareFusion, sent an "URGENT: Medical Device Recall Notification" letter dated November 8, 2013 to all customers who purchased the SmartSite Low Sorbing Infusion Set with Texium Closed Male Luer. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to immediately complete and return to CareFusion the enclosed, pre-addressed and postage paid, Recall Response Card and return all recalled product directly to the distributor form whom it was purchased. Customers with questions are instructed to contact their distributor directly or call the CareFusion Support Center at (888) 562-6018. If you have recall related questions or need support, please call CareFusion Support Center at 1-888-562-6018; hours: 7am to 4pm PST or Technical Support at 1-888-812-3229; hours 7am to 5pm PST.
Quantity
480 units
Distribution
US Distribution in CA, WY, OH, UT, and MA.
Lot, serial or model codes
Lot No:  13046215 13046641 13056285 13056577 13075931
510(k) numbers
["K944320"]
PMA numbers
not on file
Product code
FPA
Firm FEI
2016493
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register