Beckman Coulter Inc.: Access Immunoassay Systems Estradiol, Part Number: 33540 Intended use: The Access Estradiol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of Estradiol levels in human serum and plasma using the Access Immunoassay systems.

Recall number
Z-0448-2011
Recalling firm
Beckman Coulter Inc.
Product
Access Immunoassay Systems Estradiol, Part Number: 33540 Intended use: The Access Estradiol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of Estradiol levels in human serum and plasma using the Access Immunoassay systems.
Status
Terminated
Initiated
2010-03-04
Posted
2010-11-23
Terminated
2012-05-07
Reason
The recall was initiated after Beckman Coulter confirmed that three lots of Access Estradiol reagent (REF 33540) contains an incorrect raw material which may produce incorrect patient results.
Root cause
Other
Action
A Product Corrective Action (PCA) letter dated March 17, 2010, with attached PCA Customer Response Form was sent to the affected customers. The letter describes the product, problem and actions to be taken by the customers. The customers were instructed to discontinue use of the three Access Estradiol Reagent lots; discard all remaining inventory of affected product; share this information with laboratory staff and retain the notification as part of their laboratory Quality System documentation, and complete and return the enclosed response form within 10 days via fax to 786-639-4000 or mail: Beckman Coulter, Inc., 1000 Lake Hazeltine Drive, Chaska, MN 55318, Attn: Regulatory Affairs. The following product replacement contacts were provided: In the United States, please contact Client Services at 1-800-526-3821. In Canada, please contact Customer Service at 1-800-463-7828. Outside the United States and Canada, contact your local Beckman Coulter Representative. Questions and assistance could be obtained through Technical Support at 1-800-854-3633 in the United States and Canada. Outside the United States and Canada, contact consignees were told to contact their local Beckman Coulter Representative.
Quantity
20,209 units (5475 in US)
Distribution
Worldwide distribution: USA and Internationally to: Algeria, Argentina, Armenia, Australia, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Czech Republic, El Salvador, France, Georgia, Germany, Ghana, Greece, India, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Lebanon, Libyan Arab Jamahiriya, Malaysia, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nigeria, Oman, Panama, Philippines, Poland, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab, United Kingdom, Venezuela, Vietnam, and Zambia
Lot, serial or model codes
Lot Numbers: 913016 (exp. 6/20/2010), 915171 (exp. 10/30/2010), 917766 (exp. 12/21/2010)
510(k) numbers
["K973743"]
PMA numbers
not on file
Product code
CHP
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register