FDA Medical Device Recalls (openFDA)
Elekta, Inc.: ELEKTA AXESSE, REF XRT 2181, XRT 27891; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
- Recall number
- Z-0447-2024
- Recalling firm
- Elekta, Inc.
- Product
- ELEKTA AXESSE, REF XRT 2181, XRT 27891; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
- Status
- Open, Classified
- Initiated
- 2023-10-25
- Posted
- 2023-12-01
- Terminated
- not on file
- Reason
- Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
- Root cause
- Under Investigation by firm
- Action
- Elekta released Field Safety Notice to its consignees on the 10/25/2023 via email. The notice explained the issue, potential risk, and provides information for the safe use of the instrument until Elekta Field Service Representatives can perform the correction.
- Quantity
- 47 units
- Distribution
- worldwide
- Lot, serial or model codes
- UDI/DI N/A, Serial Numbers: 152140, 152617, 152691, 152004, 152683, 152994, 153935, 152620, 152809, 152854, 152867, 152885, 152964, 152974, 153192, 153217, 153527, 152499, 153668, 152519, 153118, 153102, 153204, 153166, 152017, 152112, 152325, 152749, 152754, 152755, 152756, 152860, 153006, 151874, 152538, 152839, 152921, 153149, 153267, 152736, 151793, 151802, 151812, 151960, 152600, 152678, 152838
- 510(k) numbers
- ["K210500"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 1037831
- Firm city
- Atlanta
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28