Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: ELEKTA AXESSE, REF XRT 2181, XRT 27891; Medical Charged-Particle Radiation Therapy System Accelerator, Linear

Recall number
Z-0447-2024
Recalling firm
Elekta, Inc.
Product
ELEKTA AXESSE, REF XRT 2181, XRT 27891; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
Status
Open, Classified
Initiated
2023-10-25
Posted
2023-12-01
Terminated
not on file
Reason
Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
Root cause
Under Investigation by firm
Action
Elekta released Field Safety Notice to its consignees on the 10/25/2023 via email. The notice explained the issue, potential risk, and provides information for the safe use of the instrument until Elekta Field Service Representatives can perform the correction.
Quantity
47 units
Distribution
worldwide
Lot, serial or model codes
UDI/DI N/A, Serial Numbers: 152140, 152617, 152691, 152004, 152683, 152994, 153935, 152620, 152809, 152854, 152867, 152885, 152964, 152974, 153192, 153217, 153527, 152499, 153668, 152519, 153118, 153102, 153204, 153166, 152017, 152112, 152325, 152749, 152754, 152755, 152756, 152860, 153006, 151874, 152538, 152839, 152921, 153149, 153267, 152736, 151793, 151802, 151812, 151960, 152600, 152678, 152838
510(k) numbers
["K210500"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1037831
Firm city
Atlanta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register