Skip to the content

FDA Medical Device Recalls (openFDA)

Intuitive Surgical, Inc.: 8MM, CURVED BIPOLAR DISSECTOR; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.

Recall number
Z-0447-2015
Recalling firm
Intuitive Surgical, Inc.
Product
8MM, CURVED BIPOLAR DISSECTOR; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
Status
Terminated
Initiated
2014-11-03
Posted
2014-11-25
Terminated
2015-02-03
Reason
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
Root cause
Labeling design
Action
Urgent Medical Device Recall letters were sent to affected sites on November 3, 2014. Letters inform customers of the reason for recall, risk to health and the affected instrument. A table of all of the affected IS4000 instruments affected is attached to the letter indicating the part number and product name for each. Letters also inform customers that all instruments shipped since July 2014 are manufactured with a new material and not subject to recall. Customers are to take the following actions: 1. Ensure that all affected personnel are aware of the information. 2. Identify affected product in their possession 3. Contact ISI rep for replacement. 4. Complete and return the attached Acknowledgement From. Until replacement instruments are received, the old instruments may be used provided they do not show signs of degradation and that the reprocessing instructions are followed. Questions to 800-876-1310 Option 3 ( 6 am to 5 pm PST) or ics@intusurg.com.
Quantity
509 total units, all affected devices
Distribution
US Distribution -- FL, TX, NY, NV, OR, IN, UT, and MA.
Lot, serial or model codes
Part number; 470344-04; All 8mm da Vinci Xi (IS4000) Surgical System instruments manufactured with white colored housings.
510(k) numbers
["K131861"]
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28