Natus Medical Incorporated: The product has the Catalogue/Part number 60010. Olympic Cool-Cap System is a selective head cooling system for treatment of hypoxic-ischemic encephalopathy (HIE) in infants. A cooling water cap is placed on the infant's head and the body is warmed using radiant warmers. The Cool-Cap provides 72 hours of monitored cooling. The device is labeled in parts: "***OLYMPIC COOL-CAP Cooling Unit***Natus Medical Incorporated 5900 First Avenue South ***Made in USA***Control Unit Model No. 60010***Software in this product is***Technical Service***"

Recall number
Z-0447-2013
Recalling firm
Natus Medical Incorporated
Product
The product has the Catalogue/Part number 60010. Olympic Cool-Cap System is a selective head cooling system for treatment of hypoxic-ischemic encephalopathy (HIE) in infants. A cooling water cap is placed on the infant's head and the body is warmed using radiant warmers. The Cool-Cap provides 72 hours of monitored cooling. The device is labeled in parts: "***OLYMPIC COOL-CAP Cooling Unit***Natus Medical Incorporated 5900 First Avenue South ***Made in USA***Control Unit Model No. 60010***Software in this product is***Technical Service***"
Status
Terminated
Initiated
2012-05-09
Posted
2012-11-30
Terminated
2015-01-27
Reason
The Olympic Cool-Cap's Control Module has experienced a frozen screen during use. When this occurs, the on-screen information remains on display, but the system is no longer providing cooling treatment to the infants. The visible clock displayed at the upper right hand corner of the screen stops advancing.
Root cause
Nonconforming Material/Component
Action
The firm, Natus Medical Incorporated, sent its customers an "URGENT: MEDICAL DEVICE" letter dated May 16, 2012 with an attached Field Safety Notice. The letter describes the product, problem and actions taken. Customers can call Natus Technical Service at 888-496-2887 for any questions about this recall.
Quantity
324 units
Distribution
Worldwide distribution: USA (nationwide) including states of: AK, AR, AZ, CA, CO, DE, FL, GA, HI, ID, IL, IN, KY, MA, MI, MN, MS, ND, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, WA, Puerto Rico, and countries of: Argentina, Australia, Austria, Bahrain, Canada, Chile, China, Colombia, Ecuador, Egypt, Germany, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Italy, Kuwait, Latvia, Malaysia, Mexico, Morocco, Panama, Puerto Rico, Poland, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, and Vietnam.
Lot, serial or model codes
1008; 1009; 1011; 1012; 1014; 1016; 1017; 1018; 1020; 1021; 1022; 1023; 1024; 1025; 1027; 1028; 1029; 1030; 1031; 1032; 1034; 1035; 1036; 1038; 1044; 1045; 1048; 1050; 1051; 1052; 1053; 1054; 1055; 1056; 1059; 1060; 1062; 1063; 1064; 1065; 1066; 1070; 1071; 1072; 1073; 1074; 1075; 1076; 1077; 1079; 1080; 1081; 1082; 1083; 1084; 1085; 1086; 1087; 1088; 1089; 1090; 1091; 1092; 1093; 1094; 1095; 1096; 1097; 1098; 1101; 1116; 1117; 1118; 1119; 1120; 1121; 1122; 1124; 1125; 1126; 1127; 1128; 1129; 1130; 1131; 1132; 1134; 1136; 1137; 1138; 1139; 1141; 7002; 7003; 7004; 7005; 7007; 7008; 7009; 7010; 7011; 7012; 7013; 7014; 7015; 7016; 7017; 7018; 7019; 7020; 7021; 7022; 7023; 7024; 7025; 7026; 7027; 7028; 7029; 7030; 7031; 7032; 7033; 7034; 7036; 7037; 7038; 7039; 7040; 7041; 7042; 7043; 7044; 7045; 7046; 7047; 7048; 7049; 7050; 7051; 7052; 7053; 7054; 7055; 7250; 7251; 7252; 7253; 7254; 7255; 7256; 7257; 7258; 7259; 7260; 7261; 7262; 7263; 7264; 7266; 7268; 7269; 7274; 7276; 7277; 7281; 7286; 7287; 7289; 7293; 7305; 7306; 7307; 7308; 7310; 7311; 7327; 7330; 7348; 7351; 7356; 7358; 7360; 7363; 7364; 7365; 7366; 7367; 7368; 7369; 7370; 7371; 7372; 7374; 7375; 7376; 7377; 7378; 7380; 7381; 7383; 7384; 7385; 7386; 7387; 7388; 7389; 7394; 7400; 7401; 7402; 7405; 7406; 7407; 7409; 7410; 7411; 7412; 7413; 7414; 7416; 7418; 7420; 7421; 7424; 7431; 7433; 7434; 7435; 7436; 7437; 7438; 7440; 7441; 7265; 7267; 7270; 7271; 7272; 7273; 7275; 7278; 7279; 7280; 7282; 7283; 7284; 7285; 7288; 7290; 7291; 7292; 7294; 7295; 7296; 7297; 7298; 7299; 7300; 7301; 7302; 7303; 7304; 7309; 7312; 7313; 7314; 7315; 7316; 7317; 7318; 7319; 7320; 7321; 7322; 7323; 7324; 7325; 7326; 7328; 7329; 7331; 7332; 7333; 7334; 7335; 7336; 7337; 7338; 7339; 7340; 7341; 7342; 7343; 7344; 7345; 7346; 7347; 7349; 7350; 7352; 7353; 7354; 7355; 7357; 7359; 7361; 7362; 7379; 7382; 7391; 7392; 7393; 7395; 7397; 7398; 7399; 7403; 7404; 7408; 7422; 7423; 7429; 7432.
510(k) numbers
not on file
PMA numbers
["P040025"]
Product code
MXM
Firm FEI
3018859
Firm city
Seattle
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register