Medline Industries, LP: Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; ENT DYNJ909128I; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D

Recall number
Z-0445-2026
Recalling firm
Medline Industries, LP
Product
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; ENT DYNJ909128I; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D
Status
Open, Classified
Initiated
2025-10-10
Posted
2025-11-13
Terminated
not on file
Reason
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Root cause
Nonconforming Material/Component
Action
Safety notice was provided to customers on 10/10/2025. Follow-up "URGENT MEDICAL DEVICE RECALL" notification dated 10/21/2025 was sent out on 10/21/2025. Notification was through email and first-class mail. Customers were instructed to Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product and destroy affected kits on hand. No products are to be returned to Medline. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-25-215-FG Recall Code: If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. An expanded "URGENT MEDICAL DEVICE RECALL" notification dated 10/24/2025 was provided to customers on 10/24/25, which included additional affected products. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity
96,406 total units
Distribution
International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
Lot, serial or model codes
ENT DYNJ909128I UDI-DI 40198459333355 (case), 10198459333354 (ea) Lots 25GBO685; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D UDI-DI 40198459055301 (case), 10198459055300 (ea) Lots 25GBN602
510(k) numbers
not on file
PMA numbers
not on file
Product code
OGR
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register