Parks Medical Electronics, Inc: Devices includes a Multi Port Inflator or without Multi Port Inflator (MPI). All the devices are vascular blood flow machines. Devices are distributed with pressure cuffs of varying sizes as accessories. Affected models include *2100-SX, 3000 series, 2017, 2016, 2015, 2014, 1059, 1059-A, *1059-C, *1058-C, 4000 series, 1100 series, 1060 series, 1080 series, 234 and 246. (* = current production model) All the devices are used in the study and diagnosis of vascular disease and blockages in the upper and lower extremities.
- Recall number
- Z-0443-2013
- Recalling firm
- Parks Medical Electronics, Inc
- Product
- Devices includes a Multi Port Inflator or without Multi Port Inflator (MPI). All the devices are vascular blood flow machines. Devices are distributed with pressure cuffs of varying sizes as accessories. Affected models include *2100-SX, 3000 series, 2017, 2016, 2015, 2014, 1059, 1059-A, *1059-C, *1058-C, 4000 series, 1100 series, 1060 series, 1080 series, 234 and 246. (* = current production model) All the devices are used in the study and diagnosis of vascular disease and blockages in the upper and lower extremities.
- Status
- Terminated
- Initiated
- 2012-08-08
- Posted
- 2012-12-10
- Terminated
- 2016-08-19
- Reason
- Parks Medical Electronics, Inc. was made aware of two incidents where an air supply tube from their PARKS PVR unit was inadvertently connected to an IV in the patient's arm instead of into the blood pressure cuff.
- Root cause
- Employee error
- Action
- The firm, Parks Medical Electronics, Inc., sent an "IMPORTANT SAFETY ISSUE SERIOUS HEALTH CONSEQUENCE" letter, dated September 10 2012 to their consignees/customers. The letter described the product, problem and actions to be taken. The customers are advised to order LUER UPGRADE KIT WITH MPI if the Parks device includes a Multi Port Inflator (MPI), and/or order the LUER UPGRADE KIT WITHOUT MPI the Parks device does not have a Multi Port Inflator. While waiting for the kit or its installation, consignees/customers should use extreme caution when hooking up the air supply tubes to the cuff's Luer fittings. For ordering your kit, please call 800-547-6427 option 0. For responses, questions or concerns, customer should contact Parks Medical Electronics at 888-356-9522 (Phone), 503-591-9753 (Fax) or send email to support@parksmed.com.
- Quantity
- 349 units (Domestic: 270 units - Foreign: 79 units)
- Distribution
- Worldwide distribution: USA (nationwide) and countries of: Australia, Brazil, Canada, Chile, Egypt, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Korea, Mexico, Peru, Republica Argentina, Spain, Switzerland, and Venezuela.
- Lot, serial or model codes
- Domestic Distribution: 013053, 013104, 013107, 013124, 013154, 013182, 013193, 014203, 014237, 014248, 014249, 014252, 014253, 014254, 014269, 014276, 014310, 014335, 014339, 014361, 014366, 014376, 014380, 014381, 014382, 014384, 014397, 014401, 014406, 014416, 014424, 014429, 014431, 014443, 014444, 014448, 014454, 014458, 014530, 014532, 014533, 014540, 014542, 014545, 014547, 014551, 014552, 014552, 014553, 014598, 014601, 014604, 014608, 014616, 014654, 014657, 014659, 014660, 014661, 014663, 103077, 105X-600001024, 105X-600001025, 105X-600001026, 105X-600001027, 105X-800001019, 105X-800001022, 105X-800001023, 105X-800001024, 105X-800001026, 105X-800001027, 105X-800001027, 105X-800001028, 105X-800001029, 105X-800001030, 105X-800001031, 105X-E00001001, 105X-E00001002, 105X-E00001003, 105X-E00001004, 105X-E00001005, 105X-E00001007, 105X-E00001008, 105X-E00001009, 105X-E00001010, 105X-E00001011, 105X-E00001012, 105X-E00001013, 105X-E00001014, 105X-E00001016, 105X-E00001017, 105X-E00001018, 105X-E00001019, 105X-E00001020, 105X-F00001000, 105X-F00001001, 105X-F00001002, 105X-F00001003, 105X-F00001004, 105X-F00001005, 105X-F00001005, 105X-F00001006, 105X-F00001007, 105X-F00001008, 105X-F00001009, 105X-F00001010, 105X-F00001011, 105X-F00001012, 105X-F00001013, 105X-F00001014, 105X-F00001015, 105X-F00001016, 105X-F00001017, 105X-F00001018, 105X-F00001020, 105X-F00001021, 105X-F00001022, 105X-F00001023. There are also devices that pre-dated bar codes, as well as devices with unknown codes that the firm is unable to identify. Foreign distribution: 01301, 013043, 013044, 013045, 013046, 013072, 013076, 013175, 013181, 013183, 013188, 014215, 014386, 014392, 014393, 014456, 014535, 014550, 014551, 014561, 014562, 014563, 105X-800001025, 105X-922001017, 105X-922001018, 105X-922001019, 105X-922001021, 105X-9221020, 105X-923001001, 105X-923001002, 105X-923001003, 105X-923001004, 105X-923001005, 105X-923001006, 105X-923001007, 105X-923001007, 105X-923001009, 105X-923001010, 105X-923001011, 105X-E00001000, 105X-E00001003, 105X-E00001006, 105X-E00001015, 105X-F00001019. There are also devices that pre-dated bar codes, as well as devices with unknown codes that the firm is unable to identify.
- 510(k) numbers
- ["K944495"]
- PMA numbers
- not on file
- Product code
- DPW
- Firm FEI
- 3020115
- Firm city
- Aloha
- Firm state
- OR
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28