Beckman Coulter Inc.: SYNCHRON Systems Direct LDL Cholesterol (LDLD) Reagent, REF: 969706

Recall number
Z-0441-2026
Recalling firm
Beckman Coulter Inc.
Product
SYNCHRON Systems Direct LDL Cholesterol (LDLD) Reagent, REF: 969706
Status
Open, Classified
Initiated
2025-09-17
Posted
2025-11-06
Terminated
not on file
Reason
Beckman Coulter has identified that there is an increased probability Synchron Systems Direct LDL Cholesterol (LDLD REF 969706) reagent lot M404166 may generate false high results. The issue was discovered internally and confirmed on August 19, 2025 by Beckman Coulter.
Root cause
Nonconforming Material/Component
Action
Beckman Coulter notified consignees via letter on about 09/17/2025. Consignees were instructed to discontinue use of SYNCHRON Systems LDLD reagent lot M404166 and contact Customer Technical Support Center for assistance, review the Safety Data Sheet (SDS) and discard according to laboratory s Standard Operating Procedures and/or your local regulations. The SDS is available at www.beckmancoulter.com. Additionally, Beckman Coulter recommends sharing the content of this letter with laboratory and/or medical director regarding the need to perform a retrospective review of the patient results generated from the SYNCHRON Systems LDLD reagent lot M404166 to determine if retesting is necessary.
Quantity
2,932 units
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, CA, FL, GA, HI, IA, ID, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, PR, QC, SC, TX, WA, WV and the country of Canada.
Lot, serial or model codes
REF: 969706. Lot: M404166. Expiration Date: 30Nov2025
510(k) numbers
not on file
PMA numbers
not on file
Product code
MRR
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register