Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Navigator HD l3F/15F x 28 cm Ureteral Access Sheath Set Product No M0062502270, RDF Catalog No 250-227, Sterile. Indicated for use in endoscopic procedures to facilitate the passage of endoscopes, urological instruments and for the injection of fluids into the urinary tract.

Recall number
Z-0440-2014
Recalling firm
Boston Scientific Corporation
Product
Navigator HD l3F/15F x 28 cm Ureteral Access Sheath Set Product No M0062502270, RDF Catalog No 250-227, Sterile. Indicated for use in endoscopic procedures to facilitate the passage of endoscopes, urological instruments and for the injection of fluids into the urinary tract.
Status
Terminated
Initiated
2013-11-05
Posted
2013-12-04
Terminated
2014-12-08
Reason
The 13F/15F Navigator HD Ureteral Access SheathSet are being recalled after receipt of 5 reports where the tip of the dilator has separated fro the body of the dilator.
Root cause
Component design/selection
Action
Urgent recall notification letters were sent via Federal Express Priority mail on November 5, 2013. The notification instructs the account to check their inventory, remove any affected product from their inventory, and return it to Boston Scientific. The notification also requests the completion of a Reply Verification Tracking Form as a method of documenting the presence or absence of affected product in their inventory. Follow-up communications will be made to the accounts that do not respond to this inquiry.
Quantity
274
Distribution
Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Ireland, Denmark, Italy, France, Canada, Great Britain, Switzerland, South Africa, Germany, and the Netherlands.
Lot, serial or model codes
UPN M0062502270, Catalog 250-227, Lot 15708431, 15708432, 15708433, 15812875, 15836355, 15855545, 15876213 , 15896950, 15935564, 16028474, 16044445, and 16046280
510(k) numbers
not on file
PMA numbers
not on file
Product code
FED
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register