CardioQuip, LLC: CardioQuip Modular Cooler-Heater, Model:MCH-1000(i), Power: 115V-60hz,16A

Recall number
Z-0439-2024
Recalling firm
CardioQuip, LLC
Product
CardioQuip Modular Cooler-Heater, Model:MCH-1000(i), Power: 115V-60hz,16A
Status
Open, Classified
Initiated
2023-10-19
Posted
2023-12-01
Terminated
not on file
Reason
There is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components.
Root cause
Device Design
Action
On 10/19/23, CardioQuip issued a "Urgent: Medical Device Correction" notification to all affected consignees. This notification was an update to the firm's May 22, 2023 communication. CardioQuip ask consignees to take the following actions: 1. Prime the MCH-1000(i) and MCH-1000(m) prior to each use. 2.Never operate the device dry (see communication for further details). 3. Design Changes to be implemented (see communication for further details). A. Relocation of the water Level Sensor (see communication for further details). B. Software Update (see communication for further details). 4. Please fill out and submit the receipt of understanding of the contents of this letter to CardioQuip within 5 business days or receipt. 5. Any adverse events experienced relating to CardioQuip devices should be reported to CardioQuip or through the FDA s MedWatch Program: Phone: +1 (800) FDA-1088 Web: www.fda.gov/medwatch/report.htm
Quantity
125 units
Distribution
US Distribution: TX, IN, IL, WI, FL, NJ, PA, KY, VA, WV, CT, CA, WA, AL, IA
Lot, serial or model codes
UDI: (01)00860000846103/ Serial Numbers: Serial Numbers: 10160461, 10160463, 10160465, 10160466, 101060467, 10160468, 10161406, 10161408, 10161553, 10161689, 10160275 1016276, 1016277, 10160274, 10161065, 10161066, 10160825, 10160826, 10160827, 10160828, 10161140, 10161141, 10160830, 10160831, 10160832, 10160833, 10161543, 10160449, 10160450, 10160451, 10160452 , 1016050512, 10160513, 10160781, 10160785, 10160786, 10160787, 10160788, 10160789, 10160790, 10160791, 10160792, 10160793, 10160794, 10160795, 10161130, 10161131, 10161329, 10161517, 10160758, 10160094, 10160095, 10160096, 10160110, 10160113, 10160114, 10161294, 10161301, 10161303, 10161304, 10161306, 10160570, 10160572, 10160623 ,10160797, 10160798, 10160279, 10160280, 10160281, 10160282, 10161021, 10160797, 10161199, 10161200, 10161275, 10161343, 10161344, 10161560, 10161561, 10160824, 10160767, 10160097, 10160098, 10160099, 10160104, 10160105, 10160106, 10160108, 10160109, 10161043, 10161187, 10161679, 10161680, 10161681, 10160983, 10161830, 10160517, 10160519, 10160520, 10161312, 10160190, 10160191, 10160192, 10160193, 10160194, 10160248, 10160249, 10160250, 10160251, 10160967, 10160683, 10160684, 10160685, 10160686, 10160687, 10160688, 10160689, 10160690, 10161358, 10161359, 10161362, 10160084, 10160161, 10160183, 10161841
510(k) numbers
["K102147"]
PMA numbers
not on file
Product code
DWC
Firm FEI
3007899424
Firm city
College Station
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register