Boston Scientific Scimed: Mach 1 6F Kimny Guide Catheter. Catalog No. 34356-583. Sterilized with ethylene oxide gas. Nonpyrogenic. Intended for one procedure use only. Manufacturer: Boston Scientific Scimed, Inc., One Scimed Place, Maple Grove, MN USA 55311.

Recall number
Z-0432-06
Recalling firm
Boston Scientific Scimed
Product
Mach 1 6F Kimny Guide Catheter. Catalog No. 34356-583. Sterilized with ethylene oxide gas. Nonpyrogenic. Intended for one procedure use only. Manufacturer: Boston Scientific Scimed, Inc., One Scimed Place, Maple Grove, MN USA 55311.
Status
Terminated
Initiated
2004-05-24
Posted
2006-01-24
Terminated
2007-01-23
Reason
Some Mach 1 6F Coronary Guide Catheter could be mislabeled in that a few units labeled as Mach 1 6F Kimny Curve style, could contain a Mach 1 6F JR 3.5 curve style and a few units labeled as Mach 1 6F JR 3.5 curve style, could contain a Mach 1 6F Kimny curve style.
Root cause
Other
Action
A Recall Notice, dated 05/28/05, was sent to Risk Managers. The notice informed them of the issue and instructed them to discontinue use and segregate affected units. A Tracking/verification Form is requested to be returned. Arrangements will be made for returning affected product.
Quantity
10 devices
Distribution
Within the US to GA, MS, NC ,TN. OUS to include: Canada, Hong Kong, Italy, Netherlands, Saudi Arabia, United Kingdom
Lot, serial or model codes
Product UPN H749343565830, Lot # 6287514
510(k) numbers
["K010874"]
PMA numbers
not on file
Product code
DQY
Firm FEI
3002095335
Firm city
Maple Grove
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register