FDA Medical Device Recalls (openFDA)
Stryker Orthobiologics, Inc.: Styker Orthobiologics Vita Plasma Separator, a component of Vitagel Surgical Hemostat. Sterile, single use device designed to prepare plasma from the patient's blood using a tabletop centrifuge.
- Recall number
- Z-0431-2013
- Recalling firm
- Stryker Orthobiologics, Inc.
- Product
- Styker Orthobiologics Vita Plasma Separator, a component of Vitagel Surgical Hemostat. Sterile, single use device designed to prepare plasma from the patient's blood using a tabletop centrifuge.
- Status
- Terminated
- Initiated
- 2012-07-26
- Posted
- 2012-11-22
- Terminated
- 2014-10-29
- Reason
- When preparing plasma using VitaPrep, it was reported that a small amount of blood collected in the bottom of the centrifuge cup following the centrifugation step. Firm has since discovered that if blood is present in the Luer connection of the VitaPrep following the blood draw step, this blood could be displaced to the outside of the syringe when the VitaPrep cap is reattached to the syringe. T
- Root cause
- Device Design
- Action
- Stryker sent a "MEDICAL DEVICE FIELD CORRECTION" dated July 26, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The letter included a Business Reply Form for customers to complete and return via fax to 1-484-323-8803. Contact the firm at 1-888-774-8870 for questions regarding this notice.
- Quantity
- 3,268
- Distribution
- Nationwide Distribution-including the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, PA, SC, SD, TN, TX, VA, WA, WI, and WV
- Lot, serial or model codes
- Part number 2113-0013, lot numbers T1203002, T1204005, and T1206001.
- 510(k) numbers
- not on file
- PMA numbers
- ["P050044"]
- Product code
- LMF
- Firm FEI
- 1000307073
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28