FDA Medical Device Recalls (openFDA)
Biomerieux Inc: 0.9% Sodium Chloride Injections, USP, 1000ml, VWR/Baxter Product Usage: The concentrated sodium chloride is used to prepare inoculum that is analyzed by the VITEK 2 instrument. The VITEK 2 user manual instructs all users to prepare inoculum with 0.45%-0.50% sodium chloride injections, and the system is designed to use 0.45%-0.50% sodium chloride concentration only.
- Recall number
- Z-0430-2013
- Recalling firm
- Biomerieux Inc
- Product
- 0.9% Sodium Chloride Injections, USP, 1000ml, VWR/Baxter Product Usage: The concentrated sodium chloride is used to prepare inoculum that is analyzed by the VITEK 2 instrument. The VITEK 2 user manual instructs all users to prepare inoculum with 0.45%-0.50% sodium chloride injections, and the system is designed to use 0.45%-0.50% sodium chloride concentration only.
- Status
- Terminated
- Initiated
- 2012-10-17
- Posted
- 2012-11-22
- Terminated
- 2013-12-05
- Reason
- The firm inventoried and distributed 0.9% Sodium Chloride Saline bags as 0.45% Sodium Chloride Saline bags for use with the VITEK 2 system.
- Root cause
- Employee error
- Action
- Biomerieux sent an Urgent Product Removal Notice letter dated October 16, 2012 to all affected customers via certified mail. The letter identified the affected product, problem and actions to be taken. Customers were instructed to discontinue use of the affected product and immediately discard any remaining inventory. The letter included instructions on receiving product replacement or credit. For questions or concerns contact your local BioMerieux representative.
- Quantity
- 199 cases/14-1000mL bags
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of: AL, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, PA, RI, TN, TX, VA and WI. Distribution was also made to Canada, military facilities and other foriegn countries.
- Lot, serial or model codes
- Lot number C866715, EXP OCT 13
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- not on file
- Firm FEI
- 1950204
- Firm city
- Hazelwood
- Firm state
- MO
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28