XTANT Medical: XTANT Medical, ## mm x ## mm x # mm x # ,Lordotic/Packaged InTice - C, Sterile R, Single Use Only, for the following Reference numbers/ Primary DI Number/Material Grade on Label: Ref: X098-0106-STR DI Number: M697X0980106STR1 Mat.: CP Ti Grade 3, Ref: X098-0107-STR DI Number: M697X0980107STR1 Mat.: CP Ti Grade 4, Ref: X098-0108-STR DI Number: M697X0980108STR1 Mat.: CP Ti Grade 5, Ref: X098-0109-STR DI Number: M697X0980109STR1 Mat.: CP Ti Grade 6, Ref: X098-0110-STR DI Number: M697X0980110STR1 Mat.: CP Ti Grade 7, Ref: X098-0111-STR DI Number: M697X0980111STR1 Mat.: CP Ti Grade 8, Ref: X098-0112-STR DI Number: M697X0980112STR1 Mat.: CP Ti Grade 9, Ref: X098-0205-STR DI Number: M697X0980205STR1 Mat.: CP Ti Grade 10, Ref: X098-0206-STR DI Number: M697X0980206STR1 Mat.: CP Ti Grade 11, Ref: X098-0207-STR DI Number: M697X0980207STR1 Mat.: CP Ti Grade 12, Ref: X098-0208-STR DI Number: M697X0980208STR1 Mat.: CP Ti Grade 13, Ref: X098-0209-STR DI Number: M697X0980209STR1 Mat.: CP Ti Grade 14, Ref: X098-0210-STR DI Number: M697X0980210STR1 Mat.: CP Ti Grade 15, Ref: X098-0211-STR DI Number: M697X0980211STR1 Mat.: CP Ti Grade 16, Ref: X098-0212-STR DI Number: M697X0980212STR1 Mat.: CP Ti Grade 17, Ref: X098-0305-STR DI Number: M697X0980305STR1 Mat.: CP Ti Grade 18, Ref: X098-0306-STR DI Number: M697X0980306STR1 Mat.: CP Ti Grade 19, Ref: X098-0307-STR DI Number: M697X0980307STR1 Mat.: CP Ti Grade 20, Ref: X098-0308-STR DI Number: M697X0980308STR1 Mat.: CP Ti Grade 21, Ref: X098-0309-STR DI Number: M697X0980309STR1 Mat.: CP Ti Grade 22, Ref: X098-0310-STR DI Number: M697X0980310STR1 Mat.: CP Ti Grade 23, Ref: X098-0311-STR DI Number: M697X0980311STR1 Mat.: CP Ti Grade 24, Ref: X098-0312-STR DI Number: M697X0980312STR1 Mat.: CP Ti Grade 25

Recall number
Z-0429-2021
Recalling firm
XTANT Medical
Product
XTANT Medical, ## mm x ## mm x # mm x # ,Lordotic/Packaged InTice - C, Sterile R, Single Use Only, for the following Reference numbers/ Primary DI Number/Material Grade on Label: Ref: X098-0106-STR DI Number: M697X0980106STR1 Mat.: CP Ti Grade 3, Ref: X098-0107-STR DI Number: M697X0980107STR1 Mat.: CP Ti Grade 4, Ref: X098-0108-STR DI Number: M697X0980108STR1 Mat.: CP Ti Grade 5, Ref: X098-0109-STR DI Number: M697X0980109STR1 Mat.: CP Ti Grade 6, Ref: X098-0110-STR DI Number: M697X0980110STR1 Mat.: CP Ti Grade 7, Ref: X098-0111-STR DI Number: M697X0980111STR1 Mat.: CP Ti Grade 8, Ref: X098-0112-STR DI Number: M697X0980112STR1 Mat.: CP Ti Grade 9, Ref: X098-0205-STR DI Number: M697X0980205STR1 Mat.: CP Ti Grade 10, Ref: X098-0206-STR DI Number: M697X0980206STR1 Mat.: CP Ti Grade 11, Ref: X098-0207-STR DI Number: M697X0980207STR1 Mat.: CP Ti Grade 12, Ref: X098-0208-STR DI Number: M697X0980208STR1 Mat.: CP Ti Grade 13, Ref: X098-0209-STR DI Number: M697X0980209STR1 Mat.: CP Ti Grade 14, Ref: X098-0210-STR DI Number: M697X0980210STR1 Mat.: CP Ti Grade 15, Ref: X098-0211-STR DI Number: M697X0980211STR1 Mat.: CP Ti Grade 16, Ref: X098-0212-STR DI Number: M697X0980212STR1 Mat.: CP Ti Grade 17, Ref: X098-0305-STR DI Number: M697X0980305STR1 Mat.: CP Ti Grade 18, Ref: X098-0306-STR DI Number: M697X0980306STR1 Mat.: CP Ti Grade 19, Ref: X098-0307-STR DI Number: M697X0980307STR1 Mat.: CP Ti Grade 20, Ref: X098-0308-STR DI Number: M697X0980308STR1 Mat.: CP Ti Grade 21, Ref: X098-0309-STR DI Number: M697X0980309STR1 Mat.: CP Ti Grade 22, Ref: X098-0310-STR DI Number: M697X0980310STR1 Mat.: CP Ti Grade 23, Ref: X098-0311-STR DI Number: M697X0980311STR1 Mat.: CP Ti Grade 24, Ref: X098-0312-STR DI Number: M697X0980312STR1 Mat.: CP Ti Grade 25
Status
Terminated
Initiated
2020-09-30
Posted
not on file
Terminated
2021-05-21
Reason
Cervical Implants contain a label that incorrectly identifies the grade of titanium used.
Root cause
Labeling mix-ups
Action
On 9/30/2020 and 10/29/2020, a "Urgent: Medical Device Recall" notification was sent to consignees via FedEx Priority Overnight. In addition to notifying the consignee about the labeling issue, the recall notification as consignees to take the following actions: 1. Immediately examine your inventory and quarantine any affected products subject to this recall. See enclosed illustration for ease in identifying affected products (see Attachment 2). 2. Please complete the enclosed Acknowledgement Form (Attachment 1), even if you have no remaining devices in your inventory, and return the completed Acknowledgment Form to the Attention of Michele Dolan via: Mail: Xtant Medical InTice-C Porous Ti Cervical Interbody System Recall 664 Cruiser Lane Belgrade, MT 59714 Email: mdolan@xtantmedical.com Fax:1-406-388-3380 3. Return Affected Product Contact Michele Dolan at mdolan@xtantmedical.com, or 1-406-388-0480 ext. 1127, between 8 a.m. and 5 p.m. (Mountain Time) to obtain instructions for returning affected product, and to obtain replacement devices which will be provided at no cost to you. 4. Package the affected products to be returned. Xtant will provide a shipping container (if needed) with a return shipping label. Securely package the affected product(s). Send the container at your earliest opportunity to: ATTN: InTice-C Porous Ti Cervical Interbody System Recall RA# 20-xxxx Xtant Medical 732 Cruiser Lane Belgrade, MT 59714 If you have any questions, please contact Rebecca Lennemann, Director of RAQA, 1-406-388-0480 ext. 1125, from 8 a.m. to 5 p.m. (Mountain Time). For consignees that have already implanted the device: 1. The product does not pose any health risk. The product is correctly manufactured with commercially pure Titanium (CP-Ti), Grade 2, and as such, there is no need for surgical intervention. 2. This notice is being provided only as a notification of the recall and requires completion of the Customer Acknowl
Quantity
508 units
Distribution
US: FL, TX, CA, OH, AZ, OH OUS: None
Lot, serial or model codes
Reference Number/Lot Number/Expiration date: X098-0106-STR/ 1220190000207152/ 12-13-24, X098-0107-STR / 1220190000207153/ 12-06-24,  X098-0108-STR / 1220190000207154/ 12-06-24, X098-0109-STR / 0120200000207155 / 01-10-25, X098-0110-STR / 0120200000207156 / 01-10-25, X098-0111-STR / 0120200000207157 / 01-10-25,  X098-0112-STR / 1220190000207158 / 12-07-24, X098-0205-STR / 1220190000207159 / 12-06-24, X098-0206-STR / 12201900002071510 / 12-06-24, X098-0207-STR / 12201900002071511 / 12-07-24, X098-0208-STR / 12201900002071512 / 12-07-24, X098-0209-STR / 12201900002071513 / 12-06-24, X098-0210-STR / 12201900002071514 / 12-07-24,  X098-0211-STR / 12201900002071515 / 12-07-24, X098-0212-STR / 12201900002071516 / 12-07-24, X098-0305-STR / 02202000002071517 / 02-05-25 X098-0306-STR / 12201900002071518 / 12/13/24, X098-0307-STR / 01202000002071519 / 01-10-25,  X098-0308-STR / 12201900002071520 / 1213-24, X098-0309-STR / 12201900002071521 / 12-07-24, X098-0310-STR / 12201900002071522 / 12-07-24, X098-0311-STR / 12201900002071523 / / 12-07-24, X098-0312-STR / 12201900002071524 / 12-08/24
510(k) numbers
["K173832"]
PMA numbers
not on file
Product code
ODP
Firm FEI
3005031160
Firm city
Belgrade
Firm state
MT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register