Skip to the content

FDA Medical Device Recalls (openFDA)

DiaDexus, Inc: diaDexus PLAC Test ELISA Kit; Manufactured and distributed by diaDexus , Inc., South San Francisco, CA; Model Part Number: 90123 The diaDexusPLAC Test ELISA Kit is an enzyme Immunoassay for the quantitative determination of Lp-PLA2 (lipoprotein-associated phospholipase A2) in human plasma and serum, to be used in conjunction with clinical evaluation and patient risk assessment as an aid in predicting risk for coronary hear disease, and ischemic stroke associated with atherosclerosis.

Recall number
Z-0429-2011
Recalling firm
DiaDexus, Inc
Product
diaDexus PLAC Test ELISA Kit; Manufactured and distributed by diaDexus , Inc., South San Francisco, CA; Model Part Number: 90123 The diaDexusPLAC Test ELISA Kit is an enzyme Immunoassay for the quantitative determination of Lp-PLA2 (lipoprotein-associated phospholipase A2) in human plasma and serum, to be used in conjunction with clinical evaluation and patient risk assessment as an aid in predicting risk for coronary hear disease, and ischemic stroke associated with atherosclerosis.
Status
Terminated
Initiated
2010-03-15
Posted
2010-11-22
Terminated
2011-01-11
Reason
The reason for the recall is that Real Time Stability Test program showed the kit lot failed stability prior to expiration, affecting samples with results> 500 ng/ml. Results up to 500 ng/ml exhibit passing stability results and are not impacted by this issue.
Root cause
Other
Action
DiaDexus notified all customers by phone, including distributor's customers, beginning on March 15, 2010. Customers were told to re-test samples that were tested from 2/1/2010 through 2/28/2010 which have values above 500 ng/ml with a new lot of product. For questions regarding this recall call (650) 246-6477.
Quantity
239 kits
Distribution
Nationwide Distribution - US including NJ, OH, KS, IN, FL, CA, UT, OK, PA, WA and MI.
Lot, serial or model codes
Lot Number: 907090; Part Number: 90123; Product Code NOE
510(k) numbers
not on file
PMA numbers
not on file
Product code
NOE
Firm FEI
3003643666
Firm city
South San Francisco
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register