Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Spine: Stryker Thor Anterior Plating System; Non Sterile Thor Standard Long Screw 6.0 x 40MM, spinal implant component

Recall number
Z-0429-2010
Recalling firm
Stryker Spine
Product
Stryker Thor Anterior Plating System; Non Sterile Thor Standard Long Screw 6.0 x 40MM, spinal implant component
Status
Terminated
Initiated
2009-10-02
Posted
2009-11-24
Terminated
2013-01-23
Reason
Surgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break at the tip.
Root cause
Device Design
Action
Recall notification letters were sent to all Stryker branches, Hospital Risk Management and Surgeons on October 19, 2009 by Federal Express. Questions are to be directed to Tiffani Rogers, Regulatory Compliance Manager at 201-760-8206.
Quantity
202
Distribution
Nationwide
Lot, serial or model codes
Catalog number 48036040, all lot numbers
510(k) numbers
["K073437"]
PMA numbers
not on file
Product code
KWQ
Firm FEI
3004024955
Firm city
Allendale
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register