Skip to the content

FDA Medical Device Recalls (openFDA)

Innovative Magnetic Resonance Imaging Systems Inc.: IMRIS ORT100 (Part Number 110470-000) and ORT200 (Part Number 112216-00), are tables used in both the Neuro II-SE and Neuro III-SV Magnetic Resonance Imaging Systems.

Recall number
Z-0428-2013
Recalling firm
Innovative Magnetic Resonance Imaging Systems Inc.
Product
IMRIS ORT100 (Part Number 110470-000) and ORT200 (Part Number 112216-00), are tables used in both the Neuro II-SE and Neuro III-SV Magnetic Resonance Imaging Systems.
Status
Terminated
Initiated
2012-10-19
Posted
2012-11-21
Terminated
2012-12-11
Reason
IMRIS Inc. initiated a recall of their OR Tables ORT100 and ORT200. When proper procedure is not followed, the foot end of the IMRIS intra-operative OR table may be attracted to the MRI magnet as the magnet approaches the table. This poses a potential risk of injury to both the patient and operators standing near the table.
Root cause
Labeling design
Action
Consignees were sent on 10/19/12 a IMRIS "Urgent Medical Device Correction" letter dated October 16. 2012. The letter described the product and the problem involved in the recall. It also provided changes to the Operator's Manual and labeling which addresses the issue. Customers may contact 1-866-475-0525.
Quantity
23
Distribution
Nationwide Distribution, including the states of FL, GA, IN, MD,MA, MI, MN, MO, TN, UT, and VA.
Lot, serial or model codes
10002845, 10002341, 10003244, 10002794, 10003243, 10003629, 10003801, 10002654, 10002796, 10002795, 10002846, 10002847, 10003830, 10003064, 10005201,10003754, 10004513
510(k) numbers
["K061916","K071099","K083137"]
PMA numbers
not on file
Product code
LNH
Firm FEI
3003807210
Firm city
Winnipeg Canada
Firm state
Manitoba
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register