Skip to the content

FDA Medical Device Recalls (openFDA)

Taut, Inc.: Taut Intraducer Peritoneal Catheter, 12 Fr x 10 Fr x 20.3 cm; a sterile, Rx peritoneal catheter for single use, individually packaged in a Tyvek/clear pouch; Taut Inc., 2561 Kaneville Ct., Geneva, IL 60134; the individually packaged, sterilirzed catheter was sold in two configurations: a) catalog number (REF) P.I.-128: 10 catheters per box, 10 boxes per case, and b) a sterile component of the convenience kit, catalog number (REF) 50000, System One Comprehensive Lap CBDE Kit

Recall number
Z-0428-2007
Recalling firm
Taut, Inc.
Product
Taut Intraducer Peritoneal Catheter, 12 Fr x 10 Fr x 20.3 cm; a sterile, Rx peritoneal catheter for single use, individually packaged in a Tyvek/clear pouch; Taut Inc., 2561 Kaneville Ct., Geneva, IL 60134; the individually packaged, sterilirzed catheter was sold in two configurations: a) catalog number (REF) P.I.-128: 10 catheters per box, 10 boxes per case, and b) a sterile component of the convenience kit, catalog number (REF) 50000, System One Comprehensive Lap CBDE Kit
Status
Terminated
Initiated
2007-01-19
Posted
2007-02-15
Terminated
2009-08-12
Reason
Taut, Inc. has become aware that some package seals are not intact, compromising the sterility of the catheters.
Root cause
Other
Action
Taut sent Urgent Medical Device Recall letters dated 1/19/07, via UPS over night, on the same date, informing them that the two lots of catheters may have package seals that are not intact. Customers were instructed to check their stocks for the affected lots, remove any found and return all affected product immediately. The PI-128 is also a component of the System One Comprehensive Lap CBDE Kist, catalog number 50000, which is a convenience kit. The customer is being instructed to remove the PI-128, Intraducer Peritoneal Catheter contained in the System Once Comprehensive Lap CBDE Kit in the eight kit lots listed, and return the PI-128 catheters. All other components of the kit may be used. As the PI-128, Intraducer Peritoneal Catheter, can also be ordered as a replacement component, the customer is also being instructed to check all opened System One Comprehensive lap CBDE kits, regardless of lot number, to ensure that it does not contain the impacted PI-128 lots. The customers were requested to complete and return the enclosed Acknowledgment & Stock Status Form, which includes an option to obtain replacement product for the PI-128, immediately faxing it back to Linda Todd at 919-433-4986. Any questions were directed to Linda Todd at 1-800-334-9751, ext. 4951.
Quantity
5,769 catheters
Distribution
Worldwide, including USA, Guam, Australia, Iceland and Italy
Lot, serial or model codes
catalog number (REF) PI-128, lot number 04116, use by 2008-01, and lot number 05178, use by 2008-09  These two lots of catheters were sold individually and as sterile components of catalog number (REF) 50000, kit lot numbers KC05107, KC05186, KC06002, KC06025, KC06071, KC06060, KC06087, KC06116
510(k) numbers
["K960883"]
PMA numbers
not on file
Product code
GBW
Firm FEI
1419305
Firm city
Geneva
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register