FDA Medical Device Recalls (openFDA)
Hitachi Medical Systems America Inc: The AIRIS MR System is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagital, oblique, and curved cross-sectional images that display the internal structure of the head, body, and extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance and proton density, spin-lattice relaxation time (T1) and spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
- Recall number
- Z-0428-06
- Recalling firm
- Hitachi Medical Systems America Inc
- Product
- The AIRIS MR System is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagital, oblique, and curved cross-sectional images that display the internal structure of the head, body, and extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance and proton density, spin-lattice relaxation time (T1) and spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
- Status
- Terminated
- Initiated
- 2005-12-01
- Posted
- 2006-01-24
- Terminated
- 2008-10-29
- Reason
- The bolts holding the lift mechanism may be damaged causing them to fail and the table top to drop several inches.
- Root cause
- Other
- Action
- The recalling firm sent a recall/field correction letter, dated 12/1/05 to all consignees.
- Quantity
- 32 devices
- Distribution
- The devices were distributed to FL, WA, IL, TX, NC, NY, OH, CO, LA, NJ, CA, AZ, SC, AR, OK, MN, SD, OR, PA, AL, GA, MD, MS, MO, CT, WI, KY, IN, KS, MA, DE, IA, UT, TN, VA, ID, NE, ND, NV, Puerto Rico, and to St. Thomas, Virgin Islands.
- Lot, serial or model codes
- Serial numbers A305 thru A341.
- 510(k) numbers
- ["K945155"]
- PMA numbers
- not on file
- Product code
- LNH
- Firm FEI
- 1528028
- Firm city
- Twinsburg
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28