Skip to the content

FDA Medical Device Recalls (openFDA)

Hitachi Medical Systems America Inc: The AIRIS MR System is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagital, oblique, and curved cross-sectional images that display the internal structure of the head, body, and extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance and proton density, spin-lattice relaxation time (T1) and spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

Recall number
Z-0428-06
Recalling firm
Hitachi Medical Systems America Inc
Product
The AIRIS MR System is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagital, oblique, and curved cross-sectional images that display the internal structure of the head, body, and extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance and proton density, spin-lattice relaxation time (T1) and spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
Status
Terminated
Initiated
2005-12-01
Posted
2006-01-24
Terminated
2008-10-29
Reason
The bolts holding the lift mechanism may be damaged causing them to fail and the table top to drop several inches.
Root cause
Other
Action
The recalling firm sent a recall/field correction letter, dated 12/1/05 to all consignees.
Quantity
32 devices
Distribution
The devices were distributed to FL, WA, IL, TX, NC, NY, OH, CO, LA, NJ, CA, AZ, SC, AR, OK, MN, SD, OR, PA, AL, GA, MD, MS, MO, CT, WI, KY, IN, KS, MA, DE, IA, UT, TN, VA, ID, NE, ND, NV, Puerto Rico, and to St. Thomas, Virgin Islands.
Lot, serial or model codes
Serial numbers A305 thru A341.
510(k) numbers
["K945155"]
PMA numbers
not on file
Product code
LNH
Firm FEI
1528028
Firm city
Twinsburg
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register