Thoratec Corp: HeartMate 12 volt sealed lead acid Battery Clips that are components of the HeartMate II LVAS (PMA P060040) and the HeartMate XVE LVAS (PMA P920014) Systems. The HeartMate II and the HeartMate XVE LVAS are intended for use as a bridge to transplantation in cardiac transplant candidates at risk of imminent death from non-reversible left ventricular failure. The HeartMate XVE LVAS is also indicated for use in patients with New York Heart Association (NYHA) Class IIIB or IV end-stage left ventricular failure who have received optimal medical therapy for at least 45 of the last 60 days and who are not candidates for cardiac transplantation. The HeartMate II and XVE LVAS are intended for use both inside and outside the hospital, for transportation of Ventricular Assist Device (VAD) patients via ground ambulance, fixed-wing aircraft or helicopter. The devices consist of an implantable blood pump connected to an external System Controller by a percutaneous lead. The external controller is powered by AC electrical power, via the Power Module or Power Base Unit, or by battery power with a pair of rechargeable HeartMate batteries. Two batteries are used simultaneously for power. The LVAS is also designed to operate on a single battery, facilitating battery changes without interruption of VAD support.

Recall number
Z-0422-2010
Recalling firm
Thoratec Corp
Product
HeartMate 12 volt sealed lead acid Battery Clips that are components of the HeartMate II LVAS (PMA P060040) and the HeartMate XVE LVAS (PMA P920014) Systems. The HeartMate II and the HeartMate XVE LVAS are intended for use as a bridge to transplantation in cardiac transplant candidates at risk of imminent death from non-reversible left ventricular failure. The HeartMate XVE LVAS is also indicated for use in patients with New York Heart Association (NYHA) Class IIIB or IV end-stage left ventricular failure who have received optimal medical therapy for at least 45 of the last 60 days and who are not candidates for cardiac transplantation. The HeartMate II and XVE LVAS are intended for use both inside and outside the hospital, for transportation of Ventricular Assist Device (VAD) patients via ground ambulance, fixed-wing aircraft or helicopter. The devices consist of an implantable blood pump connected to an external System Controller by a percutaneous lead. The external controller is powered by AC electrical power, via the Power Module or Power Base Unit, or by battery power with a pair of rechargeable HeartMate batteries. Two batteries are used simultaneously for power. The LVAS is also designed to operate on a single battery, facilitating battery changes without interruption of VAD support.
Status
Terminated
Initiated
2009-10-14
Posted
2009-11-24
Terminated
2011-04-18
Reason
Patient reports alerted Thoratec that during the process of disconnecting power leads from the battery clips, the threaded connector of the battery clip housing was observed to be loose, and in some cases dislodged from the battery clip housing. The problem affects HeartMate 12V SLA Battery Clips including those used for the HeartMate II LVAS and HeartMate XVE systems. If not detected and handled
Root cause
Other
Action
Firm has issued a notification letter to consignees, dated October 14, 2009 asking them to return proof of receipt, and to remove all affected unused devices from stock, and asking ongoing patients to return to hospitals for inspection, replacement if needed, and retraining on the process of switching power sources and safely switching to a back-up system controller. Contact Thoratec Customer Services at 1-800-456-1477 for any questions or concerns about this issue.
Quantity
8,194 battery clips
Distribution
Worldwide - USA- AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and the District Columbia. Internationally to Austria, Belgium, Bulgaria, Canada, Czech Republic, Denmark, France, Germany, Greece, Ireland, Israel, Italy, Lebanon, Lithuania, Luxembourg, Monaco, Netherlands, Singapore, South Africa, Sweden, Switzerland, Taiwan and the UK.
Lot, serial or model codes
Lot numbers are visible on the clips: 458560206, 581080207, 581080507, 34331720107, 34404690108, 34532990109, 343317201407, 34331720207,34331720507, 34404690208, 34404690308, 34404690408, 34432810108, 34450070108, 34477720108, 34570570109, 34577730109, 34331720307.
510(k) numbers
not on file
PMA numbers
["P060040","P920014"]
Product code
DSQ
Firm FEI
2916596
Firm city
Pleasanton
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register