FDA Medical Device Recalls (openFDA)
Integrity Implants Inc.: LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Labeled as the following sizes: a. 5.5mm x 35mm; b. 5.5mm x 40mm; c. 5.5mm x 45mm; d. 5.5mm x 50mm; e. 6.5mm x 35mm; f. 6.5mm x 40mm; g. 6.5mm x 45mm; h. 6.5mm x 50mm; i. 6.5mm x 55mm; j. 7.5mm x 35mm; k. 7.5mm x 40mm; l. 7.5mm x 45mm; m. 7.5mm x 50mm; n. 7.5mm x 55mm; o. 8.5mm x 35mm; p. 8.5mm x 40mm; q. 8.5mm x 45mm; r. 8.5mm x 50mm; s.8.5mm x 55mm; t. 9.5mm x 35mm; u. 9.5mm x 40mm; v. 9.5mm x 45mm; w. 9.5mm x 50mm; x. 9.5mm x 55mm; Used in spinal fixation surgery.
- Recall number
- Z-0420-2023
- Recalling firm
- Integrity Implants Inc.
- Product
- LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Labeled as the following sizes: a. 5.5mm x 35mm; b. 5.5mm x 40mm; c. 5.5mm x 45mm; d. 5.5mm x 50mm; e. 6.5mm x 35mm; f. 6.5mm x 40mm; g. 6.5mm x 45mm; h. 6.5mm x 50mm; i. 6.5mm x 55mm; j. 7.5mm x 35mm; k. 7.5mm x 40mm; l. 7.5mm x 45mm; m. 7.5mm x 50mm; n. 7.5mm x 55mm; o. 8.5mm x 35mm; p. 8.5mm x 40mm; q. 8.5mm x 45mm; r. 8.5mm x 50mm; s.8.5mm x 55mm; t. 9.5mm x 35mm; u. 9.5mm x 40mm; v. 9.5mm x 45mm; w. 9.5mm x 50mm; x. 9.5mm x 55mm; Used in spinal fixation surgery.
- Status
- Terminated
- Initiated
- 2022-01-27
- Posted
- 2022-12-05
- Terminated
- 2023-10-23
- Reason
- Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.
- Root cause
- Device Design
- Action
- Integrity Implants notified consignees on 01/28/2022, via emailed letter. The letter instructed consignees to quarantine affected product, complete and return the Product Tracking & Verification Form, and return affected product. On 02/05/2022 the recall was expanded, and consignees were notified via emailed letter. The listed the additional affected product and contained the same instructions to consignees.
- Quantity
- 1,369 screws
- Distribution
- Nationwide distribution to AZ, CA, CT, DC, FL, IL, IN, LA, MA, MN, NJ, OK, PR, TN, TX, VA.
- Lot, serial or model codes
- a. 5.5mm x 35mm, Part Number: LS-N6MG5535, UDI: (01)00818613028898(10)02210093, lot Number: 2210093; UDI: (01)00818613028898(10)10200053, Lot Number: 10200053. b. 5.5mm x 40mm, Part Number: LS-N6MG5540, UDI: (01)00818613028904(10)02210094, lot Number: 2210094; UDI: (01)00818613028904(10)10200054, Lot Number: 10200054. c. 5.5mm x 45mm, Part Number: LS-N6MG5545, UDI: (01)00818613028911(10)02210095, lot Number: 2210095; UDI: (01)00818613028911(10)10200055, Lot Number: 10200055. d. 5.5mm x 50mm, Part Number: LS-N6MG5550, UDI: (01)00818613028928(10)02210096, lot Number: 2210096; UDI: (01)00818613028928(10)10200056, Lot Number: 10200056. e. 6.5mm x 35mm, Part Number: LS-N6MG6535, UDI: (01)00818613028973(10)02210097, lot Number: 2210097; UDI: (01)00818613028973(10)10200057, Lot Number: 10200057; UDI: (01)00818613028973(10)10200057-1, Lot Number: 10200057-1. f. 6.5mm x 40mm, Part Number: LS-N6MG6540, UDI: (01)00818613028980(10)02210098, lot Number: 2210098; UDI: (01)00818613028980(10)10200058, Lot Number: 10200058; UDI: (01)00818613028980(10)10200058-1, Lot Number: 10200058-1. g. 6.5mm x 45mm, Part Number: LS-N6MG6545, UDI: (01)00818613028997(10)02210099, lot Number: 2210099; UDI: (01)00818613028997(10)10200059, Lot Number: 10200059; UDI: (01)00818613028997(10)10200059-1, Lot Number: 10200059-1. h. 6.5mm x 50mm, Part Number: LS-N6MG6550, UDI: (01)008181613029000(10)02210100, lot Number: 2210100; UDI: (01)00818613029000(10)10200060, Lot Number: 10200060; UDI: (01)00818613029000(10)10200060-1, Lot Number: 10200060-1. i. 6.5mm x 55mm, Part Number: LS-N6MG6555, UDI: (01)00818613029017(10)02210101, lot Number: 2210101; UDI: (01)00818613029017(10)10200061, Lot Number: 10200061. j. 7.5mm x 35mm, Part Number: LS-N6MG7535, UDI: (01)00818613026528(10)02210102, lot Number: 2210102; UDI: (01)00818613026528(10)10200062, Lot Number: 10200062. k. 7.5mm x 40mm, Part Number: LS-N6MG7540, UDI: (01)00818613026535(10)02210103, lot Number: 2210103; UDI: (01)00818613026535(10)10200063, Lot Number: 10200063; UDI: (01)00818613026535(10)06210309, Lot Number: 6210309. l. 7.5mm x 45mm, Part Number: LS-N6MG7545, UDI: (01)00818613026542(10)02210104, lot Number: 2210104; UDI: (01)00818613026542(10)10200064, Lot Number: 10200064; UDI: (01)00818613026542(10)10200064-1, Lot Number: 10200064-1. m. 7.5mm x 50mm, Part Number: LS-N6MG7550, UDI: (01)00818613026559(10)02210105, lot Number: 2210105; UDI: (01)00818613026559(10)10200065, Lot Number: 10200065; UDI: (01)00818613026559(10)10200065-1, Lot Number: 10200065-1. n. 7.5mm x 55mm, Part Number: LS-N6MG7555, UDI: (01)00818613026566(10)02210106, lot Number: 2210106; UDI: (01)00818613026566(10)10200066, Lot Number: 10200066. o. 8.5mm x 35mm, Part Number: LS-N6MG8535, UDI: (01)00818613026597(10)02210107, lot Number: 2210107; UDI: (01)00818613026597(10)10200067, Lot Number: 10200067. p. 8.5mm x 40mm, Part Number: LS-N6MG8540, UDI: (01)00818613026603(10)02210108, lot Number: 2210108; UDI: (01)00818613026603(10)10200068, Lot Number: 10200068. q. 8.5mm x 45mm, Part Number: LS-N6MG8545, UDI: (01)00818613026610(10)02210109, lot Number: 2210109; UDI: (01)00818613026610(10)10200069, Lot Number: 10200069. r. 8.5mm x 50mm, Part Number: LS-N6MG8550, UDI: (01)00818613026627(10)02210110, lot Number: 2210110; UDI: (01)00818613026627(10)10200070, Lot Number: 10200070. s. 8.5mm x 55mm, Part Number: LS-N6MG8555, UDI: (01)00818613026634(10)02210111, lot Number: 2210111; UDI: (01)00818613026634(10)10200071, Lot Number: 10200071. t. 9.5mm x 35mm, Part Number: LS-N6MG9535, UDI: (01)00818613026658(10)02210112, lot Number: 2210112; UDI: (01)00818613026658(10)10200072, Lot Number: 10200072. u. 9.5mm x 40mm, Part Number: LS-N6MG9540, UDI: (01)00818613026665(10)02210113, lot Number: 2210113; UDI: (01)00818613026665(10)10200073, Lot Number: 10200073. v. 9.5mm x 45mm, Part Number: LS-N6MG9545, UDI: (01)00818613026672(10)02210114, lot Number: 2210114; UDI: (01)00818613026672(10)10200074, Lot Number: 10200074. w. 9.5mm x 50mm, Part Number: LS-N6MG9550, UDI: (01)00818613026689(10)02210115, lot Number: 2210115; UDI: (01)00818613026689(10)10200075, Lot Number: 10200075. x. 9.5mm x 55mm, Part Number: LS-N6MG9555, UDI: (01)00818613026696(10)02210116, lot Number: 2210116; UDI: (01)00818613026696(10)10200076, Lot Number: 10200076.
- 510(k) numbers
- ["K203367"]
- PMA numbers
- not on file
- Product code
- NKB
- Firm FEI
- 3012797630
- Firm city
- Palm Beach Gardens
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28