Natus Medical Incorporated: The Natus Blue Light Phototherapy is also called the "neoBLUE LED Phototherapy System". The "neoBLUE mini LED Phototherapy System"

Recall number
Z-0420-2010
Recalling firm
Natus Medical Incorporated
Product
The Natus Blue Light Phototherapy is also called the "neoBLUE LED Phototherapy System". The "neoBLUE mini LED Phototherapy System"
Status
Terminated
Initiated
2009-10-01
Posted
2009-11-24
Terminated
2011-09-22
Reason
The clamp connecting the neoBLUE mini phototherapy light to the accessory pole stand can sustain stress fractures or break when excessive force is used when tightening the clamp. No injuries have been reported.
Root cause
Device Design
Action
On 10/1/09 Natus notified domestic consignees by Urgent Product Advisory Notice letter that advised them of the issue, instructed them to inspect the clamps, and included pictures of the affected clamp. Consignees are instructed to return the recalled clamps for a new replacement to be shipped by Natus. International distribution partners were notified by letter on 10/1/2009 and instructed to identify and contact their consignees. Replacements will be sent to the distribution partner for their consignees.
Quantity
2127
Distribution
Worldwide distribution: USA, Argentina, Australia, Austria, Belgium, Bermuda, Canada, Chile, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Kenya, Kuwait, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Philippines, Poland, Portugal, PR, Qatar, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, and UK
Lot, serial or model codes
Catalogue / Part Numbers: 010101 (110V, US, new); 010101R (110V, US, refurbish); 010102 (230V, UK, new); 010102R (230V, UK, refurbish); 010103 (230V, EUR, new); 010103R (230V, EUR, refurbish); 010104 (230V, EUR, new); 010104R (230V, EUR, refurbish).  AFFECTED SERIAL NUMBERS:  60011-60013, 60015, 60017, 60021-60029, 60031-60033, 60035, 60037, 60038, 60101-60120, 60124-60128, 60131-60324, 60327-60349, 60351-60403, 60405, 60407-60415, 60417-60507, 60509-60586, 60588-60606, 60608-60611, 60613-60845, 60847-61000, 61002-61293, 61295-61340, 61346-61574, 61577-61590, 62001-62026, 62028-62032, 62035-62145, 62147-62163, 62165-62245, 62247-62252, 62254-62309, 62311-62350, 62352-62433, 62435-62481, 62485-62634, 62636-62640, and 62642-62656.
510(k) numbers
["K022196"]
PMA numbers
not on file
Product code
LBI
Firm FEI
3018859
Firm city
Seattle
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register