Skip to the content

FDA Medical Device Recalls (openFDA)

King Systems Corp.: KING LAD SILI CONE SINGLE USE, King Systems, 10 devices/case, ETO Sterile Size 3 / Child Silicone King LAD Patient Size: 30-50kg Product Usage: Laryngeal airway

Recall number
Z-0419-2013
Recalling firm
King Systems Corp.
Product
KING LAD SILI CONE SINGLE USE, King Systems, 10 devices/case, ETO Sterile Size 3 / Child Silicone King LAD Patient Size: 30-50kg Product Usage: Laryngeal airway
Status
Terminated
Initiated
2012-11-02
Posted
2012-11-21
Terminated
2013-05-28
Reason
The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that had pink color indicators when the product packaging specified that the color should change from pink to yellow (or brown) if the product had subjected to an EtO
Root cause
Nonconforming Material/Component
Action
King Systems sent an URGENT MEDICAL DEVICE RECALL NOTIFICATION letter dated October 31, 2012 to all affected consignees via email. The letter identified affected product, problem and actions to be taken. The letter states that further use and/or distribution of any remaining product affected by this recall should cease immediately. The letter instructs customers to destroy all of the affected products and indicate the destruction by signing and returning the attached forms using the enclosed self-addressed envelope. For questions contact your King Systems customer service representative at 800-642-5464.
Quantity
263 cases of product. (2,630 devices)
Distribution
US Nationwide Distribution - WI, TX, NY, MO, IN, KY, FL, NJ, PA, TN, and NC
Lot, serial or model codes
REF: LAD-303, Lot #K120618
510(k) numbers
not on file
PMA numbers
not on file
Product code
CAE
Firm FEI
1824226
Firm city
Noblesville
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register