Skip to the content

FDA Medical Device Recalls (openFDA)

Fujifilm Medical Systems U.S.A., Inc.: Fujifilm Medical Systems Duodenoscope, Model Numbers ED-200XU, ED-200XT, ED-250XT, ED-310XU, ED-420XL, ED-410XT, ED-410XU, ED-450XL, and ED-450XT. Medical endoscope for the duodenum and upper G.I. tract. Open elevator wire channel duodenoscopes.

Recall number
Z-0418-2016
Recalling firm
Fujifilm Medical Systems U.S.A., Inc.
Product
Fujifilm Medical Systems Duodenoscope, Model Numbers ED-200XU, ED-200XT, ED-250XT, ED-310XU, ED-420XL, ED-410XT, ED-410XU, ED-450XL, and ED-450XT. Medical endoscope for the duodenum and upper G.I. tract. Open elevator wire channel duodenoscopes.
Status
Terminated
Initiated
2015-08-31
Posted
not on file
Terminated
2018-03-21
Reason
A precautionary measure because the Duodenoscopes have been redesigned to incorporate a closed elevator wire channel that facilitates reprocessing.
Root cause
Device Design
Action
FujiFilm mailed a Medical Device Recall Letter and Tracking/Verification Form dated August 31, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to check their inventory for affected product, if found, affected product should be removed from circulation and returned to Fugifilm. Customers were also asked to complete and return the enclosed tracking/verification form to Fujifilm Medical Systems USA, Inc., Endoscopy Division (FMSU-ESD), 10 High Point Drive, Wayne, NJ 07470 or via fax to 973-633-8818. Customers with questions were instructed to contact their Fujifilm sales representative. For questions regarding this recall call 973-686-2482
Quantity
79 units (US) and 985 units (Internationally)
Distribution
US including CA, CO, FL, IA, IL, KS, LA, MD, MN, MO, NC, TN and WA and Internationally to Asian Pacific, EU, Latin America, and Japan.
Lot, serial or model codes
Model Number: ED-200XU: Serial Numbers: 6572042, 5511033 and H451107  Model Number: ED-200XT: Serial Number: 8681034  Model Number: ED-250XT: Serial Number: HD077A008  Model Number: ED-310XU: Serial Number: 8591013  Model Number: ED-420XL: Serial Numbers: 9711002, 2D071D016, 2D071D010, 2D071D029 and 2D071D026  Model Number: ED-410XT: Serial Numbers: 8691039, 8691034 and 7691022  Model Number: ED-410XU: Serial Numbers: 2D060D010, 2D060D003, 6602091 and 6602100  Model Number: ED-450XL: Serial Numbers: 6D075B023, 6D075B015, 1D075B029 and 6D075B020  Model Number: ED-450XT: Serial Numbers; 1D076A002, 1D076A011, 1D076A013, 1D076A043, 1D076A039, 6D076A020, 6D076A033, 6D076A039, 6D076A037, 6D076A030 and 6D076A051
510(k) numbers
["K944759"]
PMA numbers
not on file
Product code
FDT
Firm FEI
2431293
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register