Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Healthcare Inc.: IntelliVue Clinical Information Portfolio (ICIP) Critical Care Release D, Version D.02. Intended for use in data collection, storage, and management with independent bedside devices, and ancillary systems that are connected either directly or through networks. It is indicated for use by healthcare providers whenever there is a need for generation of a patient record and computation of drug dosage.

Recall number
Z-0418-2010
Recalling firm
Philips Healthcare Inc.
Product
IntelliVue Clinical Information Portfolio (ICIP) Critical Care Release D, Version D.02. Intended for use in data collection, storage, and management with independent bedside devices, and ancillary systems that are connected either directly or through networks. It is indicated for use by healthcare providers whenever there is a need for generation of a patient record and computation of drug dosage.
Status
Terminated
Initiated
2009-10-21
Posted
2009-11-24
Terminated
2016-08-22
Reason
A software defect exists in the ICIP Release 0.02 scheduling function for order management such that if a user edits either the Schedule Time or the Stop Time for an existing order, the corresponding intervention(s) for the order may not be created.
Root cause
Software design
Action
Philips Medical Systems notified accounts by letter on October 21, 2009, of the release ICIP version D.03, which contains a software correction . A Philips representative will schedule the implementation of this upgrade. Please contact the Philips Customer Care Center with questions or concerns about this correction at 1-800-722-9377.
Quantity
53 units
Distribution
MD, RI, WV, and TX Australia Belgium Denmark France Italy Malaysia Singapore Spain Switzerland Taiwan United Arab Emirates United Kingdom and Ireland
Lot, serial or model codes
S/W Version D.02
510(k) numbers
["K992636"]
PMA numbers
not on file
Product code
DXJ
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register