FDA Medical Device Recalls (openFDA)
AVID Medical, Inc.: Halyard MINOR PACK - Medical convenience kits Model Number: MMOK013-03
- Recall number
- Z-0417-2025
- Recalling firm
- AVID Medical, Inc.
- Product
- Halyard MINOR PACK - Medical convenience kits Model Number: MMOK013-03
- Status
- Open, Classified
- Initiated
- 2024-09-20
- Posted
- 2024-11-14
- Terminated
- not on file
- Reason
- Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
- Root cause
- Process control
- Action
- Avid Medical issued Urgent: Voluntary Medical Device Recall letter via email on 9/20/24. Letter states reason for recall, health risk and action to take: If you received any of the batches/lots identified in APPENDIX 1, please review your inventory, and discard the entire kit. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. Distributors: If any of the affected kits identified in Appendix 1 were further distributed, please ensure all end-users are appropriately notified of this product correction and maintain records of effectiveness. "Complete the attached Recall Response Form and return as soon as possible. Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com.
- Quantity
- not on file
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- GTIN: 10809160411995 Lot Number: 1598132
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LRO
- Firm FEI
- 1047429
- Firm city
- Toano
- Firm state
- VA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28