Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Instruments Div. of Stryker Corporation: Cournand Modified 2-PC 19G, 5.4cm or (Guidewire Introducer Needle) 20 units per case The Stryker Guidewire Introducer Needles are designed for the percutaneous puncture of the anterior/posterior wall of a blood vessel. This provides an entry for a guidewire, vessel dilator, sheath, and catheter introduction.

Recall number
Z-0417-2016
Recalling firm
Stryker Instruments Div. of Stryker Corporation
Product
Cournand Modified 2-PC 19G, 5.4cm or (Guidewire Introducer Needle) 20 units per case The Stryker Guidewire Introducer Needles are designed for the percutaneous puncture of the anterior/posterior wall of a blood vessel. This provides an entry for a guidewire, vessel dilator, sheath, and catheter introduction.
Status
Terminated
Initiated
2015-11-23
Posted
2015-12-14
Terminated
2016-09-06
Reason
The Guidewire Introducer Needle (0910-192-000) was shipped after the documented expiration date or with minimal time until expiration. There is a potential risk of a soft tissue infection to the patient if device is used after expiration date.
Root cause
Process control
Action
Stryker sent an Urgent Medical Device Recall Notification letter dated November 24, 2015, to all affected customers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions regarding this recall, please contact Stryker Instruments: Monday-Friday 8am-5pm (EST) 269-389-2921
Quantity
120 units
Distribution
Nationwide Distribution including PA and SC.
Lot, serial or model codes
Model Number(s): 0910-192-000 Lot Numbers Affected 61406007 expiration date: 8/31/2015
510(k) numbers
["K812973"]
PMA numbers
not on file
Product code
DYB
Firm FEI
1811755
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register