Skip to the content

FDA Medical Device Recalls (openFDA)

SpineFrontier, Inc.: S-Lift Implant Guides (also known as S-LIF) Compressor Part Number: 11-50223 Product Usage: Surgical Guides allow a surgeon to place implants into tight spaces, such as a vertebral disc space, without having the implant contact the surrounding bone or tissue. In the case of S-LIFT Implant Guides, they are used to place vertebral bodies into the disc space without the body contacting the superior or inferior vertebrae.

Recall number
Z-0417-2013
Recalling firm
SpineFrontier, Inc.
Product
S-Lift Implant Guides (also known as S-LIF) Compressor Part Number: 11-50223 Product Usage: Surgical Guides allow a surgeon to place implants into tight spaces, such as a vertebral disc space, without having the implant contact the surrounding bone or tissue. In the case of S-LIFT Implant Guides, they are used to place vertebral bodies into the disc space without the body contacting the superior or inferior vertebrae.
Status
Terminated
Initiated
2012-08-22
Posted
2012-11-21
Terminated
2015-06-11
Reason
When using the optional S-LIFT (also known as S-LIF) Implant Guides without stops, there is a known risk of moving the Implant Guides, and possibly the implant, too far into the disc space creating the potential for injury to the patient. The instructions for Use for the S-Lift Implant Guides will be amended to add the following warning: "Care is to be taken to not over insert the Implant Guides
Root cause
Labeling design
Action
Spine Frontier issued a Correction Notice Letter August 22, 2012 to all affected customers to update the Instructions for Use. The letter identified the affected product, problem and actions to be taken. Customers were Instructed to remove the Instructions for Use they currently have and replace them with the update. The letter instructs customers to complete and return the Response Form to 978-232-3990.
Quantity
19 units
Distribution
US Nationwide including the states of: MA and TX.
Lot, serial or model codes
Lot Numbers: 012064-001
510(k) numbers
["K092815"]
PMA numbers
not on file
Product code
MAX
Firm FEI
3005977257
Firm city
Beverly
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register