FDA Medical Device Recalls (openFDA)
Biomet, Inc.: OSS INTERLOK BOWED IM STEM W/SCREW 14 MM X 150 MM TI 6AL 4V ALLOY. Hip and knee joint replacement prostheses intended for primary and revision joint arthroplasty for use in cemented applications.
- Recall number
- Z-0417-2012
- Recalling firm
- Biomet, Inc.
- Product
- OSS INTERLOK BOWED IM STEM W/SCREW 14 MM X 150 MM TI 6AL 4V ALLOY. Hip and knee joint replacement prostheses intended for primary and revision joint arthroplasty for use in cemented applications.
- Status
- Terminated
- Initiated
- 2011-11-01
- Posted
- 2012-01-11
- Terminated
- 2012-09-10
- Reason
- Biomet has initiated this action following an investigation which identified that Part 150367, (OSS Cemented 1M Stem 13x150) was incorrectly labeled as part 150368 (OSS Cemented 1M Stem 14x150).
- Root cause
- Mixed-up of materials/components
- Action
- Biomet, Inc. sent an "URGENT MEDICAL DEVICE RECALL NOTICE" dated November 1, 2011 to the one distributor with the affected product. The letter described the product, problem, and actions to be taken by the customer. The letter instructs the customer to immediately locate and remove the affected product from inventory. Additionally, the customer was asked to complete and return a Response Form via fax to 574-372-1683. Call 574-372-3983 for questions related to this notice.
- Quantity
- 3 units & 1 pending release
- Distribution
- Distributed only in The Netherlands.
- Lot, serial or model codes
- Item: 150368; Lot 081650
- 510(k) numbers
- ["K002757"]
- PMA numbers
- not on file
- Product code
- KRO
- Firm FEI
- 1825034
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28