FDA Medical Device Recalls (openFDA)
Cook Inc.: Guardia Access Embryo Transfer Catheter is used for transferring IVF embryos into the uterine cavity.
- Recall number
- Z-0416-2020
- Recalling firm
- Cook Inc.
- Product
- Guardia Access Embryo Transfer Catheter is used for transferring IVF embryos into the uterine cavity.
- Status
- Terminated
- Initiated
- 2019-10-03
- Posted
- not on file
- Terminated
- 2020-05-13
- Reason
- Bent transfer catheter tips may cause difficulty or inability in inserting the transfer catheter into the guiding catheter. Potential adverse events that may occur if an affected product is used include increased procedural time, the need for a repeat embryo transfer procedure, or the need for a patient to undergo an additional IVF cycle.
- Root cause
- Under Investigation by firm
- Action
- An Urgent Field Safety Notice letter dated 10/3/19 was sent to customers. The letter instructs customers to identify affected devices, quarantine the devices, return them, and complete the enclosed Customer Response Form. Where product is indicated as being returned, our Customer Services department will contact you to organize the return and issue you with the relevant Returns Authorization number. Please include contact details on the Customer Response form. Returned Product should be addressed to: Cook Medical EUDC Robert-Koch-Strae, 2 52499 Baesweiler GERMANY Credit will be provided for the returned affected products where applicable.
- Quantity
- 400
- Distribution
- No US distribution. The products were distributed to the following foreign countries: Belgium, Denmark, Germany, Italy, Netherlands, Norway, Poland, Republic of Macedonia, Slovenia, Spain, and Switzerland.
- Lot, serial or model codes
- Lot Number 8361746
- 510(k) numbers
- ["K983594"]
- PMA numbers
- not on file
- Product code
- MQF
- Firm FEI
- 1820334
- Firm city
- Bloomington
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as